10
SCC-F Sample gas feed unit Operator’s manual
42/23-51 EN Rev. 3
Installing the sample gas feed unit,
continued
Installation of the sample gas feed unit and the sample gas cooler side-by-side is
described on page 11.
Sample gas feed unit
installation on a
mounting plate
Step Action
1
Fit the unit to the mounting plate using 4 M6 screws (not supplied). The
fixing brackets required for this purpose are screwed securely on the
rear of the side panels in the factory.
To enable the cooling air to pass unobstructed out of the unit to
the rear, the projection of the fixing brackets of around 2.5 cm to
the rear wall, which is set up in the factory, must not be reduced.
Sample gas feed unit
installation in a 19-inch
cabinet/rack
Step Action
1
Unscrew the mounting brackets from the rear of the side covers and
screw them securely at the front of the side covers, flush with the front
cover, using the drill holes provided for this purpose.
2
Install the unit in a 19-inch cabinet/rack using 4 M6 screws.
Release the diaphragm
pumps transportation
restraints
Step Action
1
Use a Ph2 crosshead screwdriver to loosen the two M6x25 screws in
the base plate.
Retain the screws in case the unit needs to be transported again
in the future.