![ABB SCC-F Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/abb/scc-f/scc-f_operators-manual_2838054011.webp)
42/23-51 EN Rev. 3
SCC-F Sample gas feed unit Operator’s manual
11
Installing the sample gas feed unit and sample gas cooler
side-by-side
Installing the sample
gas feed unit and
sample gas cooler
side-by-side
The best way to install the SCC-F feed unit and the SCC-C cooler unit in a 19-inch
cabinet or a 19-inch rack is side-by-side. In this case, the sample gas feed unit
should be installed on the left and the cooler unit on the right (see Figure 2).
“Right”, “left” etc. should always be interpreted from a position facing
towards the unit’s front panel.
Figure 2
Dimensional drawing:
Sample gas feed unit
and sample gas
cooler side-by-side
(Dimensions in mm)
You must take the additional space requirement into account
adjacent to the units on the right-hand and left-hand side for the cooling air inlet
and in front of the units for the cooling air outlet (approx. 10 cm in each case),
in front of and underneath the units for running and connecting the condensate
and reagent pipes and
above the units for running and connecting the sample gas lines and the
electrical leads.
Slope max. 5°.
The user is responsible for the secure mounting of the units.
Continued on next page
26
6
30
1.
5
3
7.
8
19
0
.5
46
20
5,0
469
486
440
3
25
22
.5
Содержание SCC-F
Страница 1: ...SCC F Sample gas feed unit Operator s manual 42 23 51 EN Rev 3 ...
Страница 41: ......