N AVIG ATOR 550

  |  LO W L E V EL D I SS O LV ED OX YG EN W E T SEC T I O N  |  I N F 1 3/ 1 24 - EN R E V. B

3

6  Replacing the dissolved  

oxygen sensor

Part number: low level dissolved oxygen sensor assembly 
(AW502 080)

Removing the dissolved oxygen sensor 

assembly 

Referring to Figure 2:

Disconnect (red) sensor connector 

A

.

Unscrew clamping screw 

B

 and withdraw the sensor 

assembly from flowcell body 

C

Carefully unscrew connector nut 

G

 from sensor 

D

 and 

withdraw connector body 

E

Remove and discard O-rings 

F

 and 

H

.

Figure 2  Removing the dissolved oxygen sensor assembly