Startup
Operation
D184B133U02
FXE4000 (COPA-XE/MAG-XE)
EN - 25
Flow range end value
Nominal size
minimum (0.5 m/s)
maximum (10 m/s)
3
4
6
8
10
15
20
25
32
40
50
65
80
100
0.2 l/min
0,4 l/min
1,0 l/min
1,5 l/min
2,25 l/min
5 l/min
7,5 l/min
10 l/min
20 l/min
30 l/min
3 m
3
/h
6 m
3
/h
9 m
3
/h
12 m
3
/h
4 l/min
8 l/min
20 l/min
30 l/min
45 l/min
100 l/min
150 l/min
200 l/min
400 l/min
600 l/min
60 m
3
/h
120 m
3
/h
180 m
3
/h
240 m
3
/h
Flow range end value
Nominal size
minimum (0.5 m/s)
maximum (10 m/s)
125
150
200
250
300
350
400
450
500
600
700
800
900
1000
21 m
3
/h
30 m
3
/h
54 m
3
/h
90 m
3
/h
120 m
3
/h
165 m
3
/h
225 m
3
/h
300 m
3
/h
330 m
3
/h
480 m
3
/h
660 m
3
/h
900 m
3
/h
1200 m
3
/h
1350 m
3
/h
420 m
3
/h
600 m
3
/h
1080 m
3
/h
1800 m
3
/h
2400 m
3
/h
3300 m
3
/h
4500 m
3
/h
6000 m
3
/h
6600 m
3
/h
6900 m
3
/h
13200 m
3
/h
18000 m
3
/h
24000 m
3
/h
27000 m
3
/h
2.
Current output
(menu item “Current output”)
Select the desired current range (0 … 20 mA or 4 … 20 mA)
3. For devices with a fieldbus, the bus address must be set (menu item “Data Link”).
4.
Pulse output
(menu items “Pulse factor” and “Unit”).
To set the number of pulses per volume flow unit, a unit for the totalizer (e.g., m
3
or l) must
be selected under “Unit”. Afterward the number of pulses has to be entered in the menu item
“Pulse factor”.
5
Pulse width
(menu item “Pulse width”)
The pulse width at terminals V8 and V9 can be set between 0.1 ms and 2,000 ms.
6
System zero adjustment
(menu item “System zero adjustment”)
The fluid in the flowmeter primary must be at absolute zero. The flowmeter primary must be
full. Select the menu “System zero adjustment”. Next press ENTER. Use the STEP key to
call up “automatic” and select ENTER to start the adjustment. During the automatic
adjustment, the flowmeter primary counts from 255 to 0 in the second display line. The
system zero point adjustment is completed when the counter reaches zero. The adjustment
lasts approx. 20 seconds.