18
Sensyflow FMT200-ECO2
THERMAL MASS FLOWMETER | OI/FMT200-ECO2-EN REV. H
… 6 Commissioning
… Parameter descriptions
Device detection
Display information on the connected sensor.
Parameter
Description
Production number
Unique serial number of the connected sensor, please
provide this number with any inquiry to ABB.
Version
Sensor firmware software version.
Measured value
Figure 10: ‘Measured value’ field
In the measured value area, the current measured value is
displayed in % as a bar graph.
If the digital output of the sensor is configured as a switch
output (see
‘Switch output’ option
on page 23), the graph
displays the up-scale or down-scale of the set limit value:
• If the current measured value down-scales the set lower
limit, the bar has a
blue
color.
• If the current measured value up-scales the set upper
limit, the bar has a
red
color.
• If the current measured value is within the set limits, the
bar has a
green
color.
Standard condition
Figure 11: Standard condition field
When a volume unit is selected (e.g. Nm
3
/h), the unit always
makes reference to a defined standard condition.
In the standard condition field, the absolute pressure and
reference temperature for the selected volume unit is displayed.
When a mass flow unit is selected (e.g. kg/h) the field remains
empty.
Status bar
Display of information and help texts for the current action. In
addition, the current date and time are displayed.
Digital output
Display of the digital output configuration.
Frequency output option:
Figure 12: Frequency option
The current measured value is output as a frequency on the
digital output.
Parameter
Description
Frequency [min]
Frequency in Hz for ‘0’ flow.
Frequency [max]
Frequency in Hz for the set upper range value.