![ABB FEX300 Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/abb/fex300/fex300_operating-instruction_3745924094.webp)
Parameterization
94 FEX300,
FEX500
OI/FEX300/FEX500-EN
Menu / Parameter
Value range
Description
Device Info /
Sensor
User Span
Indication of the correction value for the sensor span
Mains Frequency
-
Mains frequency for the supply power.
Excitation Freq.
-
Frequency used to operate the magnet coils of the
flowmeter sensor.
Coil Current
-
Current used to operate the magnet coils of the
flowmeter sensor.
Pre-Amp -
Indication whether a preamplifier exists in the
flowmeter sensor or not (Yes / No).
Sensor ID
-
ID number of the sensor.
SAP / ERP No.
-
Order number of the sensor.
Term Board S/W
-
Software version of the sensor memory integrated in
the sensor.
Sensor Run Hours
-
Run hours of the flowmeter sensor.
Calibration
Select the "Calibration" submenu.
Properties
Select the "Properties" submenu.
Device Information / Sensor /
Calibration
First Cal. Date
-
Date of first calibration of sensor (calibration of new
device).
Last Cal. Date
-
Date of last calibration of sensor.
Cal. Cert. No.
-
Identification (no.) of the relevant calibration
certificate.
First Cal. Location
-
Place of first calibration of the sensor.
Last Cal. Location
-
Place of last calibration of sensor.
Cal. Mode
-
Calibration mode of the sensor.
Cal. Status
-
Calibration status of the sensor.
Device Info / Sensor /
Properties
Electrode Material
-
Electrode material of the sensor.
Lining Material
-
Liner material of the sensor.
italics
= Parameter can only be viewed at the "Advanced" password level.