FAM3200
| CI/FAM3200-EN Rev. B 17
10.2
Measuring range table
10.2.1
Model FAM3220 / FAM3250 / FAM3225 / FAM3255, variable area and armored
The data listed applies to water at 20 °C (68 °F), 1 kg/dm
3
(62.43 lb/ft
3
),1 mPas (1 cP) and to air at 0 °C (32 °F), 1013 mbar
(14.7 psia). To set up devices for different measurement media or operating conditions please check the ABB Product Selection
Assistant, available at www.abb.com/flow.
Process
connection
1)
Maximum flowrate
V
std
2)
[mPas (cp)]
P
dif
3)
[bar (psi)]
P
sta
4)
[bar (psi)]
D
p
5)
[mbar (psi)]
Order code
6)
Water
[l/h (USgal/h)]
Air
[m
3
/h (scfh)]
1/4"
1 (0.26)
35 (1.24)
4 (4)
0.7 (10.15)
1.0 (14.5) 8
(0.12) 01H
1.6 (0.42)
54 (1.91)
4 (4)
0.7 (10.15) 1.0
(14.5) 8
(0.12) 02H
2.5 (0.66)
100 (3.53)
6 (6)
0.7 (10.15)
1.0 (14.5)
8 (0.12)
03H
4 (1.06)
160 (5.65)
6 (6)
0.7 (10.15)
1.0 (14.5)
8 (0.12)
04H
6 (1.59)
230 (8.12)
18 (18)
0.7 (10.15)
1.0 (14.5)
8 (0.12)
05H
10 (2.64)
350 (12.36)
18 (18) 0.7
(10.15)
1.0
(14.5) 8
(0.12) 06H
16 (4.23)
540 (19.07)
18 (18) 0.7
(10.15)
1.0
(14.5) 8
(0.12) 07H
25 (6.6)
850 (30.02)
18 (18)
0.7 (10.15)
1.0 (14.5)
9 (0.13)
08H
40 (10.57)
1250 (44.14)
18 (18)
0.7 (10.15)
1.0 (14.5) 10.5
(0.15)
09H
60 (15.85)
1900 (67.1)
18 (18)
0.7 (10.15)
1.0 (14.5) 12.5
(0.18)
10H
100 (26.42)
3100 (109.48) 12
(12)
—
7)
1.0 (14.5)
17.0 (0.25)
11H
1) Connection nominal size
2) Vstd: Maximum permissible viscosity without calibration.
3) Pdif: Minimum required differential pressure for operation with a differential pressure regulator (FAM3220 / FAM3250 only).
4) Psta: Minimum required static pressure to prevent compression oscillations. For lower pressures, consult ABB.
5) Total pressure loss with stainless steel float at maximum flow.
6) Order code of the meter tube-float construction. See chapter Fehler! Verweisquelle konnte nicht gefunden werden. for further information.
7) Not available with differential pressure regulator.