Endura AZ25 series probe
Combustion oxygen monitor
3 Mechanical installation
12
OI/AZ25P–EN Rev. E
3.5.5 Test gas and reference air supply configurations – AutoCal systems with restrictors
Fig. 3.7 System using air as test gas 1 and
bottled gas as (optional) test gas 2
Endura AZ25 probe
Endura AZ25
AutoCal unit
Clean dry oil-free instrument air
INPUT
: 10 bar (145 psi) max.
OUTPUT
: 1 bar (15 psi) max.
5 µm / Oil
coalescing
filter regulator
Test gas 1 /
1 bar (15 psi)
Non-return
valve
Test gas 2 (optional) from bottled gas –
adjusted to 1 bar (15 psi) output
Reference air supply:
restrictor in probe limits
flow to 0.3 to 0.5 l/min
(0.64 to 1.06 scfh)
at 1 bar (15 psi)
Test gas supply:
restrictor in probe limits
flow to l/min (2.118 scfh)
at 1 bar (15 psi)
TG1
TG2
Test gas 2
Fig. 3.8 System using 2 bottled test gases
Endura AZ25 probe
Endura AZ25
AutoCal unit
Test gas 1 from bottled gas –
adjusted to 1 bar (15 psi) output
Reference
air pump
5 µm / Oil
coalescing
filter regulator
OR
Non-return
valve
Test gas 2 (optional) from bottled gas –
adjusted to 1 bar (15 psi) output
Clean dry oil-free instrument air
INPUT
: 10 bar (145 psi) max.
OUTPUT
: 1 bar (15 psi) max.
Reference air supply: restrictor
in probe limits flow to 0.3 to 0.5
l/min (0.64 to 1.06 scfh)
at 1 bar (15 psi)
Test gas supply:
restrictor in probe limits
flow to l/min (2.118 scfh)
at 1 bar (15 psi)
TG1
TG2
Test gas 2
Содержание Endura AZ25 Series
Страница 36: ...Endura AZ25 series probe Combustion oxygen monitor Notes 34 OI AZ25P EN Rev E Notes ...
Страница 37: ...Endura AZ25 series probe Combustion oxygen monitor Notes OI AZ25P EN Rev E 35 ...
Страница 38: ...Endura AZ25 series probe Combustion oxygen monitor Notes 36 OI AZ25P EN Rev E ...
Страница 39: ...Sales Software Service ...