Endura AZ25 series probe
Combustion oxygen monitor
3 Mechanical installation
14
OI/AZ25P–EN Rev. E
3.5.7 Test gas and reference air supply configurations – non-AutoCal systems without restrictors
Fig. 3.11 System using air as test gas 1 and
bottled gas as (optional) test gas 2
Endura AZ25 probe
Clean dry oil-free instrument air
INPUT
: 10 bar (145 psi) max.
OUTPUT
: 1 bar (15 psi) max.
3-Way
valve
5 µm / Oil
coalescing
filter regulator
Test gas 2 (optional) / bottled gas
adjusted to 1 bar (15 psi) output
Reference air supply:
0.3 to 0.5 l/min
(0.64 to 1.06 scfh)
ABB flowmeter adjusted to
0.3 to 0.5 l/min
(0.64 to 1.06 scfh) STP
Shut-off valve for test
gas runs >10 m (33 ft.)
ABB flowmeter
adjusted to 1 l/min
(2.118 scfh) STP
Test gas supply –
1 l/min (2.118 scfh)
Fig. 3.12 System using 2 bottled test gases
Endura AZ25 probe
Reference
air pump
3-Way
valve
5 µm / Oil
coalescing
filter regulator
Test gas 1 / bottled gas adjusted
to 1 bar (15 psi) output
Test gas 2 (optional) / bottled gas
adjusted to 1 bar (15 psi) output
Test gas supply –
1 l/min (2.118 scfh)
Shut-off valve for test
gas runs >10 m (33 ft.)
Reference air supply:
0.3 to 0.5 l/min
(0.64 to 1.06 scfh)
ABB flowmeter
adjusted to 1 l/min
(2.118 scfh) STP
ABB flowmeter
adjusted to
0.3 to 0.5 l/min
(0.64 to 1.06 scfh)
STP
Clean dry oil-free instrument air
INPUT
: 10 bar (145 psi) max.
OUTPUT
: 1 bar (15 psi) max.
OR
Содержание Endura AZ25 Series
Страница 36: ...Endura AZ25 series probe Combustion oxygen monitor Notes 34 OI AZ25P EN Rev E Notes ...
Страница 37: ...Endura AZ25 series probe Combustion oxygen monitor Notes OI AZ25P EN Rev E 35 ...
Страница 38: ...Endura AZ25 series probe Combustion oxygen monitor Notes 36 OI AZ25P EN Rev E ...
Страница 39: ...Sales Software Service ...