Endura AZ20 series probe
Combustion oxygen monitor
3 Mechanical Installation
IM/AZ20P–EN Rev. G
15
3.9.2 0.5 m (1.64 ft) Probe to Large Probe Adapter Plate
Mounting
1. Locate the adapter gasket
A
on the existing mounting
plate.
2. Locate the adapter plate
B
on the gasket.
3. Secure the gasket and plate with 6 x M10 nuts and
washers
C
.
4. Fit the probe gasket
D
over the probe end. Insert the
probe into the flue.
5. Secure the probe and gasket using 6 x M6 screws and
washers
E
.
Note.
The following procedure enables a 0.5 m (1.7 ft)
probe to be fitted into a large probe (1.0 to 4.0 m [3.3 to
13.1 ft]) ABB Standard flange hole.
Fig. 3.18 Adaptor Plate Mounting –
0.5 m (1.7 ft) Probe to Large Probe
B
C
D
A
E
Remote Probe
Arrangement
Integral Probe
Arrangement