Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
42
IM/AZ20E–EN Rev. F
…Calibrate
Cal. Diagnostics
Cal. Overdue Diag.
A diagnostic warning can be generated to advise when a
calibration is due.
Off
Off
Weeks
[1, 2, 3 or 4]
Months
[1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11 or 12]
Cal. Limits Diag.
Sets the limits for the calibration coefficients. If
calculated coefficients are outside limits, calibration fails,
the coefficients are not updated and an
Out of
Specification
diagnostic message is generated – see
Section 8.2.2, page 82. It is recommended that limits are
left at the factory default settings.
Offset
Limit
An ideal probe has an offset of 0 mV. The limit is a
deviation from the ideal – for example, if 10 mV is set,
the calibration offset has an acceptable range of –10
to 10 mV.
[0.00 to 20.00 mV]
20.00 mV
Factor
Limit
An ideal probe has a factor of 1.000. The limit is a
deviation from the ideal – for example, if 0.100 is set, the
calibration factor has an acceptable range of 0.900 to
1.100.
[0.000 to 0.100]
0.100
Cell Diagnostics
Slow
Response
Diag.
A diagnostic function to warn if the probe's response to
the injection of a test gas is slower than expected.
Disabled
Disabled
Enabled
Response
Time
Limit
Sets the maximum time allowed for a response to the
test gas. The
Test Gas Delay Time
(see page 38) is
added to this value for remote automatic calibration
systems.
[0 to 99 s]
60 s
Parameter
Comment / [Range]
Default
Table 5.7 Calibrate Menu (Continued)