Installing the Gas Analyzer
91
Fidas24: Power Supply Line Connection to 115/230 V AC
115/230-V-AC-connection cable
A connecting cable is supplied for the 115/230 V AC power supply for
heating the detector and, if applicable, the heated sample gas connection
(length 5 m, cable cross-section 3 x 1.5 mm
2
) with a 4-pole socket
connector for the connection to the analyzer module.
2
L
3
1
N
PE
The illustration shows the pin side of the plug
30
in the analyzer module
(see page 66).
The operating voltage of the detector heating is automatically detected
and switched. The set voltage can be identified through two LEDs on the
mains distribution card.
115/230 V AC Connection to an Analyzer Module
1
Attach the supplied 115/230-V-AC connecting cable with the 4-pole
socket connector to the power supply connection
30
of the analyzer
module and screw tight.
2
Connect the wires on the free end of the power cable to the power
supply.