CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
… 5 Installation
… Position and design of the gas connections
Gas connections Magnos27 (model EL3020)
Figure 26: Gas connections Magnos27 (model EL3020)
Magnos27: Gas Connections
Pos.
Connection
Supplementary information
Version
1
Pressure sensor
—
Screwed connections with hose nozzles (stainless
steel 1.4305) for hoses with 4 mm inside diameter
(included in scope of supply)
2
Not assigned
—
3
Sample gas inlet
—
4
Sample gas outlet
—
5
Purge gas inlet analyzer
—
6
Purge gas outlet analyzer
—
Note
The pressure sensor (see
on page 36) is installed as an option.
For a precise pressure correction as well as for measurements in suppressed measuring ranges, the sensor and sample gas outlet
have to be connected to each other via a T-piece and with the use of short conductors.
The lines must be as short as possible or – in the case of a greater length – have a sufficiently large inside diameter (min. 10 mm) so
that the flow effect is minimized.