![ABB EasyLine EL3000 Скачать руководство пользователя страница 172](http://html1.mh-extra.com/html/abb/easyline-el3000/easyline-el3000_operating-instruction_3689261172.webp)
172
EL3000
CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
… 12 Diagnosis / Troubleshooting
… Possible status messages
No.
Status
Message
Reaction/troubleshooting
500
iQ
An internal calibration error has occurred.
Repeat calibration. Check the configuration using Optima TCT Light
. Notify Service.
501
iQ
The requested functionality is not available in the device.
Check the configuration using Optima TCT Light .
502
iQ
A calibration error occurred in the device.
Calibration interrupted. Notify Service.
503
W S
iQ
The sensitivity is too low! The calibration was aborted.
Incorrect test gas! Check test gas supply and connection; repeat
calibration.
508
iQ
Unknown calibration error. Check configuration.
Message during automatic calibration. Check the configuration
using Optima TCT Light .
511
iQ
Auto calibration aborted.
for information
512
F
a
Autocalibration running.
for information
513
iQ
An internal calibration error has occurred.
Repeat calibration. Check the configuration using Optima TCT Light
. Notify Service.
517
F
a
Device is being serviced.
518
iQ
Calibration could not be performed because the measured value is
unstable.
Check test gas supply and connection; repeat calibration.
519
iQ
Calibration could not be performed because the preamplifier is
overranged.
Check test gas supply and connection; repeat calibration.
529
W S
iQ
The calibration was canceled, as no raw measured values can be
entered.
Check test gas supply and connection; repeat calibration.
538
W S
iQ
The zero calibration was aborted because the analyzer is dirty!
(Reference detector)
Clean sample cell. Notify Service.
539
W S
iQ
The zero calibration was aborted because the analyzer is dirty!
(measuring detector)
Clean sample cell. Notify Service.
801
A
S
a
An external failure has occurred.
Input signal at the correspondingly configured digital input.
802
W
a
An external maintenance need has occurred.
Input signal at the correspondingly configured digital input.
803
F
a
Maintenance Mode defined by the user occurred at: