46
FCM2000
D184B111U02
6.6.10 Submenu
Alarm
In the following Alarm-Menus the minimum and maximum limits for the variables mass flowrate, density and
temperature can be set. When the value is outside of the maximum or minimum setting, the contact output
can be actuated (additional settings required in the menu Contact Output). The Max-Alarm value must be
larger that the associated Min-Alarm value.
6.6.10.1
Minimum Alarm Mass Flowrate
Set the lower mass flowrate limit. It must be less than the upper mass flowrate limit.
Minimum: 0 %
Maximum: 105 %
6.6.10.2
Maximum Alarm Mass Flowrate
Set the upper mass flowrate limit. It must be greater than the lower mass flowrate limit.
Minimum: 0 %
Maximum: 105 %
6.6.10.3
Minimum Alarm Density
Set the lower density limit. It must be less than the upper density limit.
Minimum: 0.5 g/cm
3
Maximum:
3.5 g/cm
3
Submenu
Alarm
Min Alarm Qm
0.00 %
Max Alarm Qm
100.00 %
Min Alarm
Density 0.5 kg/l
6 Converter
Содержание CoriolisMaster FCM2000
Страница 6: ...Contents 6 FCM2000 D184B111U02...
Страница 90: ...11 Certificates 90 FCM2000 D184B111U02 11 Certificates 11 1 EU Certificate of Compliance Fig 42...
Страница 91: ...11 Certificates D184B111U02 FCM2000 91 Fig 43...
Страница 92: ...11 Certificates 92 FCM2000 D184B111U02 11 2 EU Type Examination Certificate Fig 44...
Страница 93: ...11 Certificates D184B111U02 FCM2000 93 Fig 45...
Страница 94: ...11 Certificates 94 FCM2000 D184B111U02 Fig 46...
Страница 95: ...11 Certificates D184B111U02 FCM2000 95 Fig 47...
Страница 96: ...11 Certificates 96 FCM2000 D184B111U02 Fig 48...
Страница 97: ...11 Certificates D184B111U02 FCM2000 97 Fig 49...
Страница 98: ...11 Certificates 98 FCM2000 D184B111U02 Fig 50...
Страница 99: ...11 Certificates D184B111U02 FCM2000 99 Fig 51...
Страница 100: ...11 Certificates 100 FCM2000 D184B111U02 11 3 PED Certificate of Compliance Fig 52...
Страница 101: ...11 Certificates D184B111U02 FCM2000 101 11 4 EHEDG Certificate Fig 53...
Страница 103: ......