Connecting the CoreSense M10
45
3
Remove the cover from the analytical unit bulkhead fitting.
4
On the attached conduit, remove the cover from over the internal cables.
5
Remove the bulkhead fitting nut and tooth washer from the small plastic bag tie-wrapped near the
conduit end.
6
From the attached conduit, route the five internal cables and conduit connector end through the
bulkhead fitting.
—
Figure 23
Cables coming from the attached conduit (once inside the cabinet)
Tooth washer
Gas lines 1 and 2
Sensor head
AC power cable
Internal
communication
cable
Analytical
unit bulkhead
fitting nut
Sensor head
DC power cable
7
Slide the tooth washer over the routed cables so that is comes in contact with the bulkhead.
8
Slide the bulkhead fitting nut over the routed cables and onto the tooth washer.
9
Screw the nut to the sensor head conduit connector head coming though to the analytical unit
bulkhead fitting.
NOTICE
To ensure that the analytical unit retains its environmental protection rating, make sure
that the cable connector is pressed hard against the fitting gasket.
Содержание CoreSense M10
Страница 1: ... INSTALL ATION GUIDE CoreSense M10 Multi gas monitoring system ...
Страница 18: ...Page intentionally left blank ...
Страница 26: ...Page intentionally left blank ...
Страница 34: ...Page intentionally left blank ...
Страница 42: ...Page intentionally left blank ...
Страница 62: ...Page intentionally left blank ...
Страница 70: ...Page intentionally left blank ...
Страница 82: ...Page intentionally left blank ...
Страница 88: ...Page intentionally left blank ...
Страница 93: ......
Страница 94: ......
Страница 95: ......