ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
61
No.
Text
S A W F Reaction/Remark
Analyzer modules
300
No new measured values from the analog/digital
converter.
x x
Notify Service.
301
The measured value exceeds the range of the
analog/digital converter.
x x
Check the sample gas concentration. Notify Service.
302
Offset drift exceeds half the permissible range.
x
Check analyzer module and sample preparation.
Permissible range: 150 % of smallest installed measurement range; 50% of physical
measurement range for Uras26. Notify service when drift exceeds these values
303
Offset drift exceeds permissible range.
x x
304
Amplification drift exceeds half the permissible
range.
x
Manually calibrate the indicated detector at zero and end point. Check analyzer
module and sample preparation.
Permissible range: 50 % the the detector sensitivity. Notify Service when drift exceeds
this value.
305
Amplification drift exceeds the permissible
range.
x x
306
The offset drift between two calibrations
exceeds the permissible range.
x
These messages are generated by automatic calibration. Check calibration for
plausibility. Fix possible cause of implausibility. Manually calibrate the indicated
detector at zero (No. 306) and end point (No. 307).
Permissible range: 15 % of the smallest installed measuring range; 6 % of the smallest
installed measuring range for measurements on plants subject to approval and
pursuant to the 27th and 30th BImSchV (Federal Regulation on Immissions)
307
The amplification drift between two calibrations
exceeds the permissible range.
x
308
A computational error occurred during the
calculation of the measured value.
x x
Notify Service.
309
The temperature regulator is defective.
x
See the status message from the applicable temperature detector
310
Temperature correction was turned off for this
component because the temperature measured
value is invalid.
x
See the status message from the applicable temperature detector
311
The pressure regulator is defective.
x x
See the status message from the applicable pressure detector
312
The pressure correction turned off for this
component because of invalid measured pressure
value.
x
See the status message from the applicable pressure detector
313
Cross-sensitivity correction is impossible for this
component because the correction value is
invalid.
x
See the status message from the applicable correction detector
314
Carrier gas correction is impossible for this
component because the correction value is
invalid.
x
See the status message from the applicable correction detector
Auxiliary detector
315
No new measured values from the analog/digital
converter.
x
Notify Service.
316
The measured value exceeds the range of the
analog/digital converter.
x
Notify Service.
317
A computational error occurred during
calculation of the measurement value.
x
Notify Service.