14 Specification
Measurement Range
Aluminium
Auto-ranging
0 to 1.500 ppm Al
Undiluted range
0 to 0.300 ppm Al
Diluted range
0.3 to 1.500 ppm Al
Ammonia
Auto-ranging
0 to 3.000 ppm NH
3
Undiluted range
0 to 0.500 ppm NH
3
Diluted range
0.5 to 3.000 ppm NH
3
Color
Range
0.00 to 500 Hazen units
0.00 to 500 mg/l Pt-Co
Iron
Auto-ranging
0 to 5.000 ppm Fe
Undiluted range
0 to 1.000 ppm Fe
Diluted range
1 to 5.000 ppm Fe
Manganese
Auto-ranging
0.02 to 10 ppm Mn
Undiluted range
0.02 to 2 ppm Mn
Diluted range
2 to 10 ppm Mn
Manganese low range
Range
0.000 to 0.100 ppm Mn
Phosphate
Auto-ranging
0.05 to 50 ppm PO
4
Undiluted range
0.05 to 10 ppm PO
4
Diluted range
10 to 50 ppm PO
4
Chemical Method
Aluminium
Pyrocatechol Violet (PCV)
Ammonia
Indophenol blue
Color
Measurement of water color by absorbance at 400nm expressed in
Alpha-Hazen scale or Platinum-Cobalt scale
Iron
Tripyridyl-triazine (TPTZ)
Manganese
Formaldoxime
Manganese low range
Malachite green
Phosphate
Phosphomolybdenum blue
Background color correction
(not applicable to Aztec 600 Color)
Compensated at the measurement wavelength
Self-cleaning
Programmable automatic chemical rinsing – piston mechanically
cleaned every measurement
Measurement Mode
Batch measurement
User-selectable measurements per hour:
Aluminium/iron/manganese: 1
to
6
Ammonia/phosphate:
1 to 4
Color:
1 to 12
Sample streams
Single or up to 3 streams – sequencing is programmable