Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
4 Installation
IM/AZT6CR–EN Rev. J
13
4.4 Mounting
1. Mark the wall using the dimensions shown in Fig. 4.2.
2. Drill and plug the three holes
A
and
B
, suitable for M6
or
1
/
4
in. screws/bolts.
3. Screw in the top bolt
A
, leaving a gap of 20 mm
(0.78 in.) between the screw head and the wall.
4. Hang the analyzer onto the top bolt.
5. Secure the analyzer to the wall using two screws
B
.
4.4.1 Reagent Mounting Tray (Option)
If used, place the reagent mounting tray no more than 1100 mm
(43.3 in) from the analyzer’s bottom plate – see Fig. 4.4,
page 14.
To secure the shelves to the wall:
1. Mark the wall using the dimensions shown in Fig. 4.3.
Alternatively, with the shelf supported carefully against the
wall, mark through the mounting holes.
2. For each tray, drill and plug mounting holes suitable for M8
or
5
/
16
in. fixings.
All Dimensions in mm (in)
Fig. 4.2 Mounting the Analyzer
Note.
Clearance – the enclosure doors can open 180°. If
mounting in a confined area, allow sufficient clearance for
cables on the door hinge side (min. 270 mm [10.6 in.]) and
100 mm (3.93 in.) on door opening side.
Note.
It is not possible to tighten this screw once the
analyzer is attached to the wall.
Hang the analyzer onto the top bolt and check that it
is retained firmly against the wall.
A
B
460
(18.1)
150 (5.9)
All Dimensions in mm (in)
Fig. 4.3 Reagent Mounting Tray (Option)
565 (22.2)
600 (23.6)
235
(9.2)
300 (11.8)
10
(0.39)
300 (11.8)
6 Holes Ø
8.5 (0.34)
35
(1.4)
35
(1.4)
400
(15.7)