51
6
INSTALLATION
…6.5
Panel-mount Analyzer Connections
6.5.2
Connections – Fig. 6.11
L
Line
N
Neutral
E
Connect supply earth (ground) to stud on case
A4
C
A5
NC
Relay 1
A6
NO
A7
C
A8
NC
Relay 2
A9
NO
A10 C
A11 NC
Relay 3 (see
Note 1
below)
A12 NO
A13 +
Analog Output 1
A14 —
A15 +
Analog Output 2
A16 —
Terminal Block B
Terminal block A
Temperature Compensator
Connections
Common
Link
TC
B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
B15
B16
Common
Link
TC
Temperature Compensator
Connections
Terminal Block C
(Option Board)
C1
Not Used
C2
Not Used
C3
C4
C5
C6
Not Used
C7
C
C8
NC
Relay 4
C9
NO
C10 C
C11 NC
Relay 5
C12 NO
C13 +
Analog Output 3
C14 —
C15 +
Analog Output 4
C16 —
Earth (Ground)
Stud on Case
(see Fig. 6.10)
Used for optional
RS485 connection.
Refer to IM/PROBUS
85 to 265 V AC +
12
or
to
Power
24 V AC
–
24V DC
Supplies
Before making any electrical connections,
see Warnings on page 45
Terminal Block B
4-Electrode Conductivity Cell Connections
Sensor B
Sensor A
B1
B9
Temperature Compensator Common
B2
B10
Link B10 to B9 (and B2 to B1 if Dual Input Analyzer)
B3
B11
Temperature Compensator
B4
B12
Screen
B5
B13
Drive –
B6
B14
Sense –
B7
B15
Sense +
B8
B16
Drive +
Fig. 6.11 Connections, Panel-mount Analyzers
Содержание AX413
Страница 68: ...66 NOTES ...