19
A: Temp.Coeff.
2.00
%/
C
A: Temp.Comp.
set to
Linear
A: Filter Time
10
Secs
Config. Sensor A
CONFIG. ALARMS
Config. Sensor B
User Temp Comp
Edit
-----
A: Temp.Comp.
set to
User
Yes
No
A: Enter Point 1
A: Filter Time
10
Secs
Config. Sensor A
CONFIG. ALARMS
Config. Sensor B
Yes
No
…5.3
Configure Sensors
Temperature Coefficient
Enter the temperature coefficient (
x 100) of the solution (0.01 to 5.0%/°C). If unknown,
the temperature coefficient (
) of the solution must be calculated – see Appendix A1.1.
If the value has not yet been calculated, set it to 2%/°C provisionally.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
configuration.
See Section 5.4.
Edit User Defined Temperature Compensation
Select
Yes
to manually define a 6-point temperature compensation curve, otherwise
select
No
.
Continued on next page.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
configuration.
See Section 5.4.
5
PROGRAMMING…