41
7
CALIBRATION…
…7.3
Factory Settings
Factory Settings Access Code
Enter the required code number, between 00000 and 19999, to gain access to the factory
settings. If an incorrect value is entered, access to subsequent parameters is prevented
and the display reverts to the top of the
Factory Settings Page
.
Calibrate Sensor A
✶
Note.
The values in the display lines for sensor calibration are shown only as
examples – the actual values obtained will differ.
Dual input conductivity only – Sensor B calibration is identical to Sensor A
calibration.
Single input conductivity only – see page 43.
Return to
Operating Page
.
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
✶
Note.
The upper 6-segment display shows the measured input voltage. Once the
signal is within range the lower 6-segment display shows the same value and
Calib
is
displayed to indicate that calibration is in progress.
Resistance Span (2k )
Set the cell simulator to 2k
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page…
Factory Set Code
0000
FACTORY SETTINGS
-----
Cal. Sensor A
-----
A:Res.Zero (O/C)
0.230
0.230
Volts
Calib
A:Res.Span (2K0)
2.015
2.015
Volts
Calib
A:Res.Zero (O/C)
0.230
0.230
Volts
Calib
A: Res.Vref
Cal. Sensor B
Cal. Output 1
Dual Cond.