SETUP
Sensor Cal. A
-----
SENSOR CAL.
-----
A: Calibration
-----
Edit
Reset
Cal. User Code
0000
A: Sensor Slope
A: Reset?
Edit
Reset
Sensor Cal. B
SENSOR CAL.
Sensor Calibration
Note.
Applicable only if
Enable Cals.
is set to
Yes
– see Section 5.3.
Sensor Calibration Security Code
Enter the required code number, between 00000 and 19999, to gain access to the sensor
calibration procedure. If an incorrect value is entered, access to subsequent calibration
pages is prevented and the display reverts to the
SENSOR CAL.
page.
Note.
Applicable only if
Alter Cal. Code
is not set to zero – see Section 5.8.
Calibrate Sensor A
Dual input conductivity only – Sensor B calibration is identical to Sensor A
calibration.
Single input conductivity only – return to main menu.
Edit or Reset Calibration
Select
Edit
to manually adjust the Slope and Offset values of the process and temperature
sensors.
Select
Reset
to reset the sensor calibration data to the standard default settings:
Sensor and Temperature Slope = 1.000
Sensor and Temperature Offset = 0.0
If
Edit
selected – continued on next page…
If
Reset
selected – continued on next page…
4.1
Sensor Calibration