The conductivity analyzer utilizes the computational ability of a
microprocessor to achieve ultra-pure water temperature
compensation using only a single platinum resistance
thermometer and mathematically calculating the temperature
compensation required to give the correct conductivity at the
reference temperature.
Figure 34 Ultra-pure water temperature compensation
µS cm
–1
Temperature °C
m
icr
o
Sie
m
en
s/
cm
Curve A – Theoretical ultra-pure water conductivity
Curve B – High purity water conductivity
(ultra-pure water with slight impurity)