2- W I R E CO N D U C T I V I T Y, P H/O R P, P I O N T R A N SM I T T ER | O I/AW T 2 10 - EN R E V. C
…6 Electrical installation
pH/ORP/pIon sensor module connections
NOTICE
ORP (Redox) and Antimony pH sensors do not feature
temperature compensation therefore do not have
temperature sensors or related wiring.
Standard sensors without diagnostic functions
NOTICE
Ensure sensor diagnostics are
Off when using standard
sensors without diagnostic functions.
Sensor type
RTD wiring
SENSE
1
GUARD
2
REF
3
S .GND
4
RTD 1
5
RTD 2
6
SHIELD
7
RTD 3
8
2867
2-lead
Clear
–
Black
–
Red
White
–
–
TB5
2-lead
Blue
–
Black
–
Red
White
–
–
AP1xx
2-lead
Clear
–
Black
–
Red
Red
White
–
–
3-lead
Clear
–
Black
–
White
Red
–
Red
AP3xx
2-lead*
Blue
–
Black
–
Red
White
–
–
3-lead
Blue
–
Black
–
Red
White
–
Grey
APS1xx
APS5xx
APS7xx
2-lead*
Blue
Yellow
Black
–
Red
White
–
–
3-lead
Blue
Yellow
Black
–
Red
White
–
Grey
* Cut and remove grey wire
Standard sensors with diagnostic functions
NOTICE
Ensure sensor diagnostics are
On when using standard
sensors with diagnostic functions.
Sensor type
RTD wiring
SENSE
1
GUARD
2
REF
3
S .GND
4
RTD 1
5
RTD 2
6
SHIELD
7
RTD 3
8
TBX5
2-lead
Blue
Yellow
Black
Green
Red
White
Dark green
–
AP2xx
2-lead*
Clear
Red
Blue
Green/Yellow
Red
White
–
–
3-lead
Clear
Red
Blue
Green/Yellow
Red
White
–
Grey
* Cut and remove grey wire
NOTICE
AWT210 pH sensor modules are supplied standardized to
theoretical sensor characteristics. Following installation,
but before use, a process calibration should be performed
to ensure optimum accuracy. For pH sensor calibration
procedures see page 42.
NOTICE
BNC adaptor option
For pH/ORP/pION sensors using a BNC connector, ABB
recommends using the optional BNC adapter.
ABB does not recommend stripping or cutting sensor
cabling due to the nature of the signal and cabling used.