50
AO2000-LS25
LASER ANALYZERS | OI/AO2000-LS25-EN REV. D
… 9 Commissioning
Setting the IP address on AO2000-LS25
Figure 34: Menu “TCP/IP”
The analyzer’s IP address is set in the “TCP/IP & modem
configuration” menu in the service program, see
modem configuration <TCP/IP & modem configuration>
page 64.
Setting up the laser analyzer in AO2000
1. Call up the “MENU / Configure / System / Setup system
modules” menu and press the <NEW> key.
• The IP address shown in the picture is an example.
2. Enter the 14-digit serial number of the analyzer (located on
the back of the transmitter housing on the name plate). The
first three digits must be “008” and the serial number is
specific to each LS25 laser analyzer.
3. Confirm the entry by pressing <ENTER>.