Hydrazine Monitor
7 Calibration
IM/7835–EN Rev. P
19
7 Calibration
7.1 Calibrating the System – General Tasks
Depending on the operating conditions, perform a calibration
every 1 to 4 weeks.
Rinse out the standard solution container with a small amount of
fresh standard solution before filling and open the outlet valve
from the container.
7.2 Sensor Calibration
To perform a sensor calibration:
1. At the transmitter, enter the
Standard
access level and, at
the
Calibrate
menu, press
. A list of calibration
parameters is displayed:
Calibrate
Sensor Calibration
Standard Solution
Restore Defaults
2. Use the
/
keys press to highlight the
Sensor
Calibration
option and press
.
The
Sensor Calibration
page is displayed.
3. Press
to start the calibration.
A bar indicates the calibration progress and the message
Settling-Please Wait
is displayed beneath the time bar.
(Sensor calibration typically lasts 15 minutes.) The
hydrazine value and process temperature are also
displayed.
4. When calibration is complete, the hydrazine value (as
µg kg) efficiency and calibration coefficient ratings are
displayed, for example:
Hyd
80.0 µg/kg
Efficiency 100.0%
Coefficient 0.1917
5. Press
to exit the
Sensor Calibration
option and return
to the
Calibrate
page.
6. To return to the
Operator
page, press
or, to set the
Standard Solution
value, press
and proceed to step 3
in Section 7.3.
7.3 Setting the Standard Solution Value
To set the
Standard Solution
value:
1. At the transmitter, enter the
Standard
access level and, at
the
Calibrate
menu, press
. A list of calibration
parameters is displayed:
Calibrate
Sensor Calibration
Standard Solution
Restore Defaults
2. Use the
/
keys press to highlight the
Standard
Solution
option and press
.
The
Calibrate
page is displayed.
3. Press
to edit the
Standard Solution
value and press
the
key to move to each number in the value.
Use the
/
keys to increase/decrease each number.
4. When value is set, press
to return to the
Calibrate
page.
5. Press
twice to return to the
Calibrate
main menu.
7.4 Restoring Calibration Defaults
To restore calibration coefficients to their factory-set defaults:
1. At the transmitter, enter the
Standard
access level and, at
the
Calibrate
menu, press
. A list of calibration
parameters is displayed:
Calibrate
Sensor Calibration
Standard Solution
Restore Defaults
2. Use the
/
keys press to highlight the
Restore
Defaults
option and press
.
The
Calibrate
page is displayed with a blank
Restore
Defaults
progress bar.
3. Press
to start the
Restore Defaults
routine.
The progress bar indicates completion status and, when
the routine is complete, the bar reverts to empty.
4. Press
twice to return to the
Calibrate
main menu.
Note.
During the calibration and recovery period the alarm
functions and analog outputs are maintained at their existing
value(s) prior to calibration.