![ABB 7835 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/abb/7835/7835_user-manual_4576018.webp)
7835
Hydrazine Monitor
6 Programming
16
IM/7835–EN Rev. P
6 Programming
6.1 Operator Page and Keys
The Operator page, icons and front panel keys are shown in Fig. 6.1:
6.1.1 Access Level and Entering Passwords
Passwords can be set to enable secure end-user access to the
Standard
level. Passwords are set, changed or restored to their
default settings at the
Device Setup/Security Setup
parameter – see page 17.
To enter the
Access Level
:
1. From the
Operator
page
(see Fig. 6.1)
, press
.
The
Access Level
is displayed:
2. Use the
/
keys to scroll to and highlight the required level.
3. Press
to select the level highlighted.
If no password has been set, the
Calibration
menu is shown – see Section 6.2.1, page 17.
If a password has been set, the
Enter Password
menu is shown and a password must be entered to access further menus –
see
Device Setup
/
Security Setup
(page 17) for details of how to setup passwords.
4. To return to the
Operator
page from
Access Level
, press
.
Fig. 6.1 Operator Page and Keys
CM30
80.0
20.0
Deg. C
Hydrazine µg/kg
In Recovery
Bargraph
Access Level Icon
Front Panel
Keys
(see below)
1
Navigation
(left)
key
2
Reserved for future
enhancements
Active Alarm Icons
Process Temperature
Hydrazine Reading
a
b
c
2
4
3
Up / Down keys – navigate up / down
menus and increase / decrease displayed
values
4
Navigation (right) key
Operator Page
Diagnostic Status Bar
Note.
When the monitor is powered-up for the first time the
Standard
level can be accessed without password protection.
Protected access to this level must be allocated on-site as required.
Menu
Access
Logout
Displayed after the
Standard
level has been accessed. Logs the user out of
Standard
level. If passwords are set, a password must be entered to access
Standard
level again after selecting
Logout
.
Read Only
Enables most parameter settings to be viewed as read-only fields.
Standard
Enables access to the
Standard
level and adjustment of all enabled parameters
for system setup, configuration and control.
Access Level
Back
Select
Standard
Read Only
Logout