7835
Hydrazine Monitor
4 Electrical Installation
IM/7835–EN Rev. P
9
4.2 Transmitter Unit – Access to Terminals
Referring to Fig. 4.2:
1. Unlock the door
A
and open fully.
2. Remove all 4 screws
B
and remove the terminal cover
plate
C
.
4.3 Transmitter Unit –
Cable Gland Entries and Connections
1. Cut the cable from the sensor unit to a suitable length to
reach the transmitter for termination on transmitter
terminal blocks.
2. Push the end of the cable through one of the supplied
glands in the gland plate.
3. Prepare the cable end and attach it to the terminal block.
Individual terminal blocks can be unplugged from the
terminal board pins if required.
4. Pass the remaining cables through the glands. Prepare
the cable ends and attach them to the terminal blocks as
shown in Fig. 4.3.
Make earth (ground) connections to the stud terminal on the
bottom of the transmitter unit.
Warning.
Isolate the transmitter from power supplies
before removing the cover.
Fig. 4.2 Transmitter Unit Access to Terminals
B
A
C
Fig. 4.3 Transmitter Unit Cable Gland Entries and Electrical Connections
SCRN
BLACK
WHITE
GREEN
RED
I OUT 2+
I OUT 2–
I OUT 1+
I OUT 1–
RANGE 2*
RANGE 1*
E
N
L
ALARM 2
N/O
C
N/C
ALARM 1
N/O
C
N/C
CAL FAIL
CAL
INDICATOR
Cable Gland Entries
*Refer to Fig.
4.4
, page
10
for details of resistor networks for remote range indication