32
…10
PROGRAMMING
Security Code
English
English
SET UP OUTPUTS
SET UP OUTPUTS
ALARM
H2 IN CO2
AIR IN CO2
Security Code
Enter the required code number, between 00000 and 19999, to gain access to the secure
parameters. If an incorrect value is entered, access to subsequent programming pages is
prevented and the display reverts to the Operating page.
Note.
The security code is preset at '00000' to allow access during commissioning
but should be altered to a unique value, known only to authorized operators – see Section
10.2.3
Set Up Outputs Page
.
Advance to
Language Page
.
10.2.2
Language Page
Advance to
Set Up Outputs Page
.
10.2.3
Set Up Outputs Page
Page Header – SET UP OUTPUTS
Advance to next parameter.
Purge Gas Alarm Select
Select the parameter on which the purge gas alarm is to operate.
Reprogram as required to agree with selected range:
i.e. for
Range 2 select %H2 IN CO2
Range 3 select %AIR IN CO2.
Continued on next page…
LOWER DISPLAY
10.2
Range 2
10.2.1
Access to Secure Parameters
A 5-digit code is used to control access to the secure parameters.