12

4 69 0  S E R I E S  |

 T U R B I D I T Y SEN S O R | O I/AT S 410 - EN R E V. A

Calibration verification with a secondary 

standard

To check the calibration:

Close the isolating valve that is installed upstream of the 
sensor.

Close the sensor inlet valve. 

Open the drain valve.  Let the sensor drain. 

NOTICE

When you remove the wiper unit, do not put too much force 
on the wiper because it can bend the wiper arm out of 90°. 
When you remove the wiper, make sure that the wiper arm is 
not bent.

Carefully remove the wiper unit (7998 011 and 012) or the 
wiper plug (7998 016 and 017) to aid complete drainage of 
the system.

When the system is empty, close the drain valve.

Thoroughly dry the flow chamber internally using clean 
tissue.

Thoroughly clean and dry the emitter and receiver lenses 
using clean tissue.

If there is condensation on the emitter and receiver lenses: 

Leave the sensor open to let the lenses reach ambient 
temperature before you do the calibration.

Insert the dry calibration standard with the zero NTU 
indication (see Figure 9 or Figure 10) facing the optical 
receiver, ensuring the locating lug engages correctly.  Refer 
to Figure 11.

10 

Note the reading on the display.

11 

Remove the dry standard, turn it through 180° and refit it, 
ensuring that the NTU value indication (refer to Figure 9 o
Figure 10) faces the receiver and the locating lug engages 
correctly.  Refer to Figure 11.

12 

Note the reading on the display.

13 

If the readings noted at steps 9 and 10 exceed ±5 % of dry 
standard value:

Repeat the procedure from step 7. 

If the readings are still outside this range, calibrate the 
sensor.

14 

Remove the dry standard and put it in its storage container.

15 

Refit the wiper unit (7998 011 and 012) or wiper plug  
(7998 016 and 017).

16 

Open the inlet valve and ensure that the flow through the 
sensor is 0.5 to 1.5 L/min

–1

.

N

U

STA

N

D.

N

U

NTU value indication

Zero NTU indication

Figure 9  Dry calibration standard for low range sensors

STA

N

D.

N

U

STA

N

D.

N

U

NTU value indication

Zero NTU indication

Figure 10  Dry calibration standard for high range sensors

NTU value 

indication

Zero NTU 

indication

Locating lug

Optical 

emitter

Optical 

receiver

Figure 11  Inserting dry standard

Содержание 4690 Series

Страница 1: ...ry standard verification and automatic cleaning The sensor is designed for use with the ABB AWT420 dual input transmitter For more information Publications for the associated transmitters are availabl...

Страница 2: ...bubbler 8 Mounting the de bubbler 8 Setup procedure for the de bubbler 9 4 Turbidity sensor electrical connections 10 5 Calibration 11 Care and maintenance of secondary standards 11 Calibration verifi...

Страница 3: ...equipment Potential safety hazards 4690 turbidity sensor electrical damage to the equipment WARNING Bodily injury To ensure safe use when operating this equipment the following points must be observe...

Страница 4: ...as far as possible The European Waste Electrical and Electronic Equipment WEEE Directive that initially came into force on August 13 2005 aims to reduce the waste arising from electrical and electron...

Страница 5: ...omponents Sample outlet Flow chamber Inlet isolating valve Drain valve Optical receiver Optical emitter Wiper unit Flow chamber Sample outlet Wiper plug Optical emitter Optical receiver Drain valve In...

Страница 6: ...ponents of each sensor Referring to Figure 2 or Figure 3 1 Install the sensor in the position shown with the brackets that are provided Make sure that the sensor is installed within 5 of its vertical...

Страница 7: ...imum flow rate of 0 5 L min to prevent solids settling in the pipework Increase the flow rate if it is necessary but do not exceed the maximum flow rate of 1 5 L min When measuring turbidity it is imp...

Страница 8: ...r Mount the de bubbler vertically with the flow upwards Disconnect fittings 12 0 5 ID tube rotatable through 360 Drain outlet Sample outlet Removable fitting 12 0 5 ID tubing 150 5 9 CRS 6 5 0 25 540...

Страница 9: ...ormance These devices are not supplied with the 4690 Turbidity systems Figure 6 Typical system installation for 4690 series turbidity systems NOTICE 1 To prevent degassing of the sample which can caus...

Страница 10: ...idity sensor electrical connections Link To AWT420 transmitter Green Red Blue Yellow Blue Screen Red Emitter Blue Green Red Screen Receiver Blue Red Screen Green Yellow Wiper unit optional Screen Whit...

Страница 11: ...ndary standards ABB s secondary standards may be used repeatedly but must be monitored for deterioration All secondary standards can change gradually with time Deterioration can be detected by measuri...

Страница 12: ...he zero NTU indication see Figure 9 or Figure 10 facing the optical receiver ensuring the locating lug engages correctly Refer to Figure 11 10 Note the reading on the display 11 Remove the dry standar...

Страница 13: ...e of the system 5 When the system is empty close the drain valve 6 Thoroughly dry the flow chamber internally using clean tissue 7 Thoroughly clean and dry the emitter and receiver lenses using clean...

Страница 14: ...ses of unstable or erratic readings Look for air bubbles in the sample The bubbles might be due to degassing of the sample caused by either a drop in sample pressure or a rise in temperature Frequent...

Страница 15: ...5 psi Ambient operating temperature 0 to 50 C 32 to 122 F Ambient operating humidity Up to 95 RH 1 Maximum measured error across full measurement range limited by uncertainty in Formazine standards 2...

Страница 16: ...Pozi Pan screw black Strategic spares Part number Description Kit contents 7998024 Feed drain kit 1 7998149 2 0216509 2 0216510 7998026 Emitter unit ISO infrared LED version 1 7998101 7998029 Receiver...

Страница 17: ...4690 SERIES TURBIDIT Y SENSOR OI ATS410 EN REV A 17 Notes...

Страница 18: ...18 4690 SERIES TURBIDIT Y SENSOR OI ATS410 EN REV A Notes...

Страница 19: ...4690 SERIES TURBIDIT Y SENSOR OI ATS410 EN REV A 19 Notes...

Страница 20: ...errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third partie...

Отзывы: