background image

4 69 0  S E R I E S  |

 T U R B I D I T Y SEN S O R | O I/AT S 410 - EN R E V. A

11

5  Calibration 

A key feature of the 4690 Series Turbidity system is the dry 
secondary calibration standard. This is designed to simplify 
routine calibration verification and minimizes the need for 
chemical standards. 

Each dry calibration standard is verified against a primary 
formazine standard before delivery, and the nominal turbidity 
value is indicated on the label.

For regulatory monitoring, use the secondary calibration 
standard for monthly calibration verification and periodically 
monitor it for deterioration using a primary standard.

Definitions 

• 

Primary standards 

 

Turbidity standards that are traceable and equivalent to 
the reference turbidity standard, within statistical errors. 
Formazine is the most commonly acceptable form of 
primary standard. The other is a commercially 
manufactured liquid suspension of styrene divinylbenzene 
polymer beads (SDB). Primary standards are used to 
calibrate a turbidity analyzer directly or to calibrate a 
secondary standard. 

• 

Secondary standards 

 

Standards that the manufacturer (or an independent 
testing organization) has certified give analyzer 
calibration results equivalent (within certain limits) to the 
results obtained when the analyzer is calibrated with a 
primary standard. 

• 

Calibration

 

A procedure that checks or adjusts an analyzer’s accuracy 
by comparison with a defined standard or reference. 

• 

Calibration verification

 

A procedure used to check whether or not the calibration 
of the analyzer is within certain limits.

Care and maintenance of secondary 

standards

ABB’s secondary standards may be used repeatedly but must 
be monitored for deterioration. 

All secondary standards can change gradually with time. 
Deterioration can be detected by measuring the turbidity value 
of the secondary standard after calibration of the analyzer with 
a primary calibration standard. 

It is recommended that secondary standards are checked every 
three months against a primary standard on the analyzer they 
are intended to be used with.

If the comparison with a primary standard shows that the 
turbidity value of the secondary standard has changed, the 
secondary standard can be assigned a new turbidity value for 
use in future calibration verification.

Included with each sensor is a calibration record card kit for 
recording the value of the secondary standard compared to the 
primary calibration data for each individual analyzer with which 
the secondary standard is used.

Additional calibration record card kits can be purchased from 
ABB (part no. 7998190 for pack of 3).

Turbidity Monitor Primary Calibration Card

Sensor Serial Number:

Dry Standard Serial Number:

Time/Date of Primary

Calibration

Dry Standard

Reading after

Primary Calibration

(NTU)

Performed By

Reorder Code: 7998190

Issue A, 08/06/2011

ABB Limited, Oldends Lane, Stonehouse, GL10 3TA

Time/Date of Primary

Calibration

Dry Standard

Reading after

Primary Calibration

(NTU)

Performed By

Figure 8  Example of calibration record card

To minimize deterioration of the secondary standard:

•   Carefully clean any residual moisture after use.
•   Clean the prism with a cloth that has no lint.
•   Do not touch the dry standard rod. Fingerprints on the 

surface can have an effect on its stated value.

•  Keep it in the container provided when not in use, and keep 

it in a dry place.

Содержание 4690 Series

Страница 1: ...ry standard verification and automatic cleaning The sensor is designed for use with the ABB AWT420 dual input transmitter For more information Publications for the associated transmitters are availabl...

Страница 2: ...bubbler 8 Mounting the de bubbler 8 Setup procedure for the de bubbler 9 4 Turbidity sensor electrical connections 10 5 Calibration 11 Care and maintenance of secondary standards 11 Calibration verifi...

Страница 3: ...equipment Potential safety hazards 4690 turbidity sensor electrical damage to the equipment WARNING Bodily injury To ensure safe use when operating this equipment the following points must be observe...

Страница 4: ...as far as possible The European Waste Electrical and Electronic Equipment WEEE Directive that initially came into force on August 13 2005 aims to reduce the waste arising from electrical and electron...

Страница 5: ...omponents Sample outlet Flow chamber Inlet isolating valve Drain valve Optical receiver Optical emitter Wiper unit Flow chamber Sample outlet Wiper plug Optical emitter Optical receiver Drain valve In...

Страница 6: ...ponents of each sensor Referring to Figure 2 or Figure 3 1 Install the sensor in the position shown with the brackets that are provided Make sure that the sensor is installed within 5 of its vertical...

Страница 7: ...imum flow rate of 0 5 L min to prevent solids settling in the pipework Increase the flow rate if it is necessary but do not exceed the maximum flow rate of 1 5 L min When measuring turbidity it is imp...

Страница 8: ...r Mount the de bubbler vertically with the flow upwards Disconnect fittings 12 0 5 ID tube rotatable through 360 Drain outlet Sample outlet Removable fitting 12 0 5 ID tubing 150 5 9 CRS 6 5 0 25 540...

Страница 9: ...ormance These devices are not supplied with the 4690 Turbidity systems Figure 6 Typical system installation for 4690 series turbidity systems NOTICE 1 To prevent degassing of the sample which can caus...

Страница 10: ...idity sensor electrical connections Link To AWT420 transmitter Green Red Blue Yellow Blue Screen Red Emitter Blue Green Red Screen Receiver Blue Red Screen Green Yellow Wiper unit optional Screen Whit...

Страница 11: ...ndary standards ABB s secondary standards may be used repeatedly but must be monitored for deterioration All secondary standards can change gradually with time Deterioration can be detected by measuri...

Страница 12: ...he zero NTU indication see Figure 9 or Figure 10 facing the optical receiver ensuring the locating lug engages correctly Refer to Figure 11 10 Note the reading on the display 11 Remove the dry standar...

Страница 13: ...e of the system 5 When the system is empty close the drain valve 6 Thoroughly dry the flow chamber internally using clean tissue 7 Thoroughly clean and dry the emitter and receiver lenses using clean...

Страница 14: ...ses of unstable or erratic readings Look for air bubbles in the sample The bubbles might be due to degassing of the sample caused by either a drop in sample pressure or a rise in temperature Frequent...

Страница 15: ...5 psi Ambient operating temperature 0 to 50 C 32 to 122 F Ambient operating humidity Up to 95 RH 1 Maximum measured error across full measurement range limited by uncertainty in Formazine standards 2...

Страница 16: ...Pozi Pan screw black Strategic spares Part number Description Kit contents 7998024 Feed drain kit 1 7998149 2 0216509 2 0216510 7998026 Emitter unit ISO infrared LED version 1 7998101 7998029 Receiver...

Страница 17: ...4690 SERIES TURBIDIT Y SENSOR OI ATS410 EN REV A 17 Notes...

Страница 18: ...18 4690 SERIES TURBIDIT Y SENSOR OI ATS410 EN REV A Notes...

Страница 19: ...4690 SERIES TURBIDIT Y SENSOR OI ATS410 EN REV A 19 Notes...

Страница 20: ...errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third partie...

Отзывы: