APPENDIX B
B1
Determination of the Correction Factor
B1.1
Definition
Suspended solids:
The weight of solid, expressed as mg l
-1
, obtained by filtration
through a 1 to 2
m filter, dried to constant weight at 105
C.
Samples not readily filtered:
The weight of solid, expressed as mg l
-1
, obtained by
centrifuging sample at 3200 G, dried to constant weight at
105
C.
B1.2
Method
Follow the calibration procedures detailed in Section 8, then
carry out the following operations:
1) On
Set Up Parameter
page select
mg/l
.
2) Enter a FTU to mg/l correction factor of 1.0.
3) Take a note of the on-line value.
4) Take a grab sample for analysis to determine the 'true'
suspended solid value.
5) Obtain the correction factor by dividing the analyzed value by
the on-line value,
e.g.
if analyzed value
=
150
and on-line value
=
100
(divide 150 by 100 = 1.5)
the correction factor is 1.5.
6) Enter the correction factor using the
and
keys.
Note.
If the instrument is configured for suspended solids,
the calibration is still carried out as turbidity, using either the
dry standard or a formazine solution.