28
mV Zero1 (0 mV)
x x x x x
ELECTRICAL CAL
— — — — —
Calibrate YES
— — — — —
NO
No
Advance to
Adjust RTX Zero
mV Span1(3000 mV)
x x x x x
mV Zero2 (0 mV)
x x x x x
x x x x x
mV Span2(4000 mV)
Adjust RTX Zero
— — — — —
Adjust RTX Span
— — — — —
Return to
top of page
Turbidity
(NTU)
0
.
6 0
Yes
…6
PROGRAMMING & ELECTRICAL CALIBRATION
6.7
Electrical Calibration Page
In this section the actual values denoted by 'xxxxx' are unimportant and are used to determine display reading stability when carrying
out the electrical calibration procedure.
Electrical Calibration
Select
YES
to access the electrical calibration sequence. Select
NO
to advance to
Adjust
RTX Zero
.
Millivolt Zero (7997 200 Series, 7997 300)
Set the millivolt source to 0 mV and allow the instrument display to stabilize.
Millivolt Span (7997 200 Series, 7997 300)
Set the millivolt source to +3000 mV and allow the instrument display to stabilize.
Millivolt Zero (7997 400 Series)
Set the millivolt source to 0 mV and allow the instrument display to stabilize.
Millivolt Span (7997 400 Series)
Set the millivolt source to +4000 mV and allow the instrument display to stabilize.
Note.
Retransmission adjustment is achieved using 4 to 20 mA. The correct values for 0 to
20 and 0 to 10 mA will then be derived automatically.
Adjust Retransmission Zero
Set the milliammeter reading to 4 mA.
Adjust Retransmission Span
Set the milliammeter reading to 20 mA.
Note.
The retransmission range selected in the
Set Up Outputs Page
does not affect the
reading.
Return to the
Operating Page
.