All sensor and sealing plugs except for the center position are secured by means of a setscrew in the side of the
top end plate. Start by unscrewing the setscrew for the wanted position (the setscrew will stop when
sufficiently extracted). Pull out the sealing plug (or sensor) by inserting a screwdriver in the slot between the
plug and the top end plate, see
Figure 5-2 Removal of sealing plug. Use a small piece of paper/cardboard to protect the lacquer of the top-
end plate
Make sure that the bore surface in the sensor mounting hole is clean and smooth and check also that the O-
rings at the sensor foot are free from dust and particles and greased with silicon grease. Align the orientation
pin in the sensor foot with the orientation hole in the top end plate, and carefully insert the sensor.
When fully seated on the top end plate, tighten the set screw with only moderate force.
Note! Always replace O-rings when connecting to a sensor or a sealing plug.
Apply Tectyl 506 (included in maintenance kit) in the slit between the Sensor and the top end plate, refer
. This will prevent crevice corrosion of the top end plate.
Figure 5-3 Apply Tectyl in the slit between the Sensor and the top end plate. (Example showing tectyl around
an electrical terminal).