23
Resetting the counter
Hold down the COUNTER switch for 4 seconds or more.
The counter data will be reset to zero.
Printing the counter graph
Press and hold the COUNTER switch, press the PAPER FEED
switch to display the current measurement count.
Total
: The total measurement count of the product.
Weekly Count : The count for last one week.
Monthly Count : The count for last one year.
Note1
:
The Total can not be reset by pressing the COUNTER switch.
Note2
:
When the print selection is "OFF", the counter graph is not printed.
9-4 Error Codes
Table 9 Error codes
Error code
Description
pe
The printer paper has run out. Install a new roll of printer paper.
HU
The printer head is up. Lower the lever and lock the printer paper in position.
err
A measurement error has occurred. Refer to “10. Troubleshooting”.
9-5 Maintenance
Do not open the device. It uses delicate electronic components and an intricate air unit that could be damaged.
If you can not fix the problem using the troubleshooting instructions, request service from your supplier or from
the A&D service group. The A&D service group will provide technical information, spare parts and units to
authorized suppliers.
The technical testing procedures, which should be done at least every two years, can be performed either by
the manufacturer or by an authorized repair service in accordance with the regulations governing
manufacturing of medical products.
9-6 Cleaning
CAUTION
Never immerse the TM-2655/TM-2655P in water for cleaning, that may damage the electronic
parts inside.
Never use thinner or a strong detergent for cleaning, that may discolor or deform the plastic
case and display panel.
Fig.20 COUNTER switch