42
Rubber pump
COUNT button
Connection hose
[
C
A
Air connector
Cap
B
Inspection equipment
500 cc air tank
Coupling connector
Cover for pressure
inspection area
1. Hold the
COUNT
button on the rear of the blood pressure monitor, and turn the
POWER
switch on.
2. “
” appears in the clock display section.
3. With “
” displayed, press the
START/STOP
button.
Pressure inspection mode starts and the current pressure is displayed.
4. Using the rubber pump, apply the pressures listed below. Compare and check the pressures
of the blood pressure monitor and the inspection equipment.
No
Pressure
setting
Instrumental error A-B
(standard)
1
0 mmHg
0 mmHg
2
50 mmHg
Within ±6 mmHg
3
200 mmHg
5. Confirm that the values are within standards. To exit the pressure inspection mode and
return to the standby mode, switch the power off and switch the power on again.
Note
Use the coupling connector for exclusive use with the TM-2657P.
A: Pressure displayed by the inspection
equipment
B: Diastolic and systolic pressures displayed
by the monitor