background image

 

 

42

 

 

Rubber pump 

COUNT button 

   

 

 
   

Connection hose 

[

 

   
 

 

 

 

Air connector

 

 
 

Cap

 

 

 
 

Inspection equipment

 

 

 

500 cc air tank 

 
 

 

Coupling connector 

 

Cover  for  pressure 
inspection area 

 

 
1.  Hold the 

COUNT 

button on the rear of the blood pressure monitor, and turn the 

POWER

 

switch on.   

2.   “

” appears in the clock display section.   

3.   With “

” displayed, press the 

START/STOP

 button. 

      Pressure inspection mode starts and the current pressure is displayed.   
4.  Using the rubber pump, apply the pressures listed below. Compare and check the pressures 

of the blood pressure monitor and the inspection equipment.   

 

No   

Pressure 
setting 

Instrumental error A-B   
(standard)   

0 mmHg   

0 mmHg   

50 mmHg   

Within ±6 mmHg   

200 mmHg   

 
5.   Confirm that the values are within standards. To exit the pressure inspection mode and 

return to the standby mode, switch the power off and switch the power on again. 

 

 

Note 

  Use the coupling connector for exclusive use with the TM-2657P.   

 

A: Pressure displayed by the inspection 

equipment 

B:  Diastolic and systolic pressures displayed 

by the monitor   

Содержание TM-2657P

Страница 1: ...TM 2657P Automatic Blood Pressure Monitor Instruction Manual 1WMPD4002893C 1906 English Original...

Страница 2: ...bed or translated into any language in any form by any means without the express written consent of A D Company Limited The contents of this manual and the specifications of the instruments covered by...

Страница 3: ...Examples The symbol indicates Caution The nature of the caution required is described inside or near the symbol using text or a picture The example on the left indicates caution against electrical sho...

Страница 4: ...ed vibrated or impacted including during transportation Avoid locations where chemicals are stored or gas occurs Avoid locations where removal and insertion of AC power cable is prohibited Installatio...

Страница 5: ...numb 3 When using the monitor Do not use a mobile phone near the monitor It may cause a malfunction Do not use the monitor in a moving vehicle as this may result in inaccurate measurements Do not use...

Страница 6: ...Label the monitor with a sign that says Out of order or Do not use and then contact A D immediately 6 When performing a maintenance inspection For your safety before performing a maintenance inspecti...

Страница 7: ...h medical electrical devices not to use mobile phones or small wireless devices High frequency noise is being introduced from other devices via the electrical outlet Check for the source of noise and...

Страница 8: ...ent is not possible Measurement is not possible with patients with peripheral hypoperfusion very low blood pressure or low body temperature since blood flow to the measurement location is low Measurem...

Страница 9: ...shipping Check the monitor for damage when unpacking it Before using the monitor ensure that everything is included and then check the main unit and each standard accessory for damage For optional it...

Страница 10: ...viii Blank page...

Страница 11: ...Attaching the instruction panel 12 6 5 Pre inspection 13 7 BLOOD PRESSURE MEASUREMENT 14 8 SETTING THE CLOCK 15 9 PRINTER 16 9 1 Installing the printer paper 16 9 2 Selecting the print format 18 10 C...

Страница 12: ...1 1 External input output unit 32 12 MAINTENANCE 41 12 1 Inspection and safety management 41 12 2 Cleaning 43 12 3 Periodic inspection 46 12 4 Replacing the arm cuff cover 47 12 5 Checking the number...

Страница 13: ...e dangerous Do not attempt to use this device on newborns or infants Using this device on small children could cause injury to them This device is designed for measuring adults Facilities with the dev...

Страница 14: ...l number Date of manufacture RS 232C serial interface EC directive medical device label WEEE label EU authorized representative Manufacturer Exx Error code display xx 00 to 99 Displays extent of elect...

Страница 15: ...appear on the display with the measurement values Note We recommend contacting your physician if you see this I H B symbol frequently When is the IHB mark printed The IHB mark is printed in the measur...

Страница 16: ...30 240 bpm NIBP Clinical test EN1060 4 2004 Pulse rate accuracy 5 Cuff Winding mechanism operated by geared motor Applicable arm circumference 18 35 cm Inflation Automatic inflation by air pump Defla...

Страница 17: ...equipment Part 1 General requirements for safety and essential performance IEC 60601 1 2 2014 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Coll...

Страница 18: ...pressure during measurement 7 Pulse display Displays the pulse measurement value 8 Clock display Displays the current time 24hour TM 2657P EX 12hour TM 2657P EG 9 Measurement status LED Displays the m...

Страница 19: ...5 Arm cuff cover Inner cover of the cuff 6 Cuff section Holds the arm cuff cover 7 Armrest Location to rest the arm during measurement 8 External input output unit The optional external input output u...

Страница 20: ...2657 03 External input output unit RS 1ch option No Name Description 1 D Sub 9 pin male RS 232C TM 2657 05 External input output unit RS Bluetooth option No Name Description Bluetooth Bluetooth Ver 2...

Страница 21: ...be relaxed To prevent theft we recommend using a chain to connect the security slot and stand See 6 3 Security slot 6 2 Power connection To avoid the risk of electric shock the monitor must only be c...

Страница 22: ...below to attach the instruction panel to the rear side of the monitor Make sure to attach the instruction panel to the main unit before use The instruction panel contains precautions that the patient...

Страница 23: ...over attached Is the arm cuff cover overstretched Connection cable Are the optional cables inserted firmly into the connectors of the monitor Power cable Make sure that the electrical outlet is proper...

Страница 24: ...ssure 5 After about one minute of measurement the cuff automatically deflates to its original state 6 The measurement results are displayed 7 The measurement results are printed on the printer paper R...

Страница 25: ...e monitor returns to standby mode Example Setting the clock to 4 37 PM April 20 2015 1 Hold the SELECT button for 1 second The systolic display section starts flashing 2 Press the button to display 20...

Страница 26: ...ver button to open the printer cover Open printer cover button Printer cover 2 Install the printer paper in the way shown in the illustration below 3 With the end of the paper at the top and protrudin...

Страница 27: ...button for 2 seconds will cut the paper Note If the direction of the printer paper is incorrect printing is not performed Use genuine A D printer paper If genuine A D paper is not used the print may b...

Страница 28: ...ilable for selection High speed printing 1 Normal 3 line printing 2 Big font printing 3 Table printing 4 For each mode mean arterial pressure MAP printing can be set to on or off F09 3 Graph printing...

Страница 29: ...at 1 EU format F27 Time format 24 24 hour F08 ID printing 3 F05 IHB on No IHB detected F11 Measurement value printing 1 High speed printing F09 MAP printing on F26 Date format 1 EU format F27 Time for...

Страница 30: ...n again Setting items Details Default Diastolic display section Function F01 Not used F02 Display time Measurement result display time seconds F03 Applied pressure Applied pressure setting mmHg F04 No...

Страница 31: ...esult input output STD RI RB BP RA D Sub RVX compatibility Mini DIN Blood pressure result input output STD RI RB BP RA D Sub A D weight scale Mini DIN Blood pressure result input output STD RI RB BP R...

Страница 32: ...ng including ID Bar code printing CODE39 with check digit modulus43 QR code printing V2 including ID F31 Bluetooth connection timing Connection at the end of measurement Connection at the start of mea...

Страница 33: ...e setting This setting appears in the diastolic display section If automatic applied pressure Aut is set pulsation is observed while pressure is applied and the applied pressure value is automatically...

Страница 34: ...ID printing can be set using the function F08 Use the button to change the setting This setting appears in the diastolic display section Only TM 2657VP TM 2657P DIA LED ID printing setting Default ID...

Страница 35: ...ppears in the diastolic display section DIA LED Mean arterial pressure printing Default Mean arterial pressure MAP printing off Mean arterial pressure MAP printing on When mean arterial pressure MAP p...

Страница 36: ...ng mode Default High speed printing Normal 3 line printing Big font printing Table printing When Mean arterial pressure MAP printing is off Printing example High speed printing Big font printing Norma...

Страница 37: ...rs in the diastolic display section DIA LED Bitmap printing Default Bitmap printing off Standard pattern printing User pattern printing For details about bitmap registration see 15 SENDING BITMAP PATT...

Страница 38: ...output STD RI RB BP RA D Sub A D weight scale Mini DIN Blood pressure result input output STD RI RB BP RA D Sub RVY compatibility External input output unit TM 2657 03 DIA LED External input output un...

Страница 39: ...e setting This setting appears in the diastolic display section DIA LED Stop bit Mini DIN Default Stop bit 1 Stop bit 2 10 16 Stop bit D Sub The stop bit D Sub can be set using the function F24 Use th...

Страница 40: ...ting appears in the diastolic display section DIA LED ICT printing Default ICT printing OFF Bar code printing CODE39 QR code printing including ID Bar code printing CODE39 with check digit modulus43 Q...

Страница 41: ...Bluetooth connection timing The Bluetooth connection timing can be set using the function F31 Use the button to change the setting This setting appears in the diastolic display section DIA LED ICT pr...

Страница 42: ...ment connected to the device must be located out of reach of the patient The personal computer or ID reader used must conform to EN60601 1 11 1 External input output unit unit function TM 2657 01 Mini...

Страница 43: ...bit 1 bit 2 bits can be changed using F23 Code ASCII Pin assignment Do not connect to Pins No 4 7 or 8 They are used for the blood pressure monitor Cable specifications for computer connection TM 2657...

Страница 44: ...nged using F20 Data bit length 8 bits Parity None Stop bit 1 bit 2 bits can be changed using F24 Code ASCII Pin assignment Pin No Signal name Description 1 2 RXD Receive data 3 TXD Transmit data 4 DTR...

Страница 45: ...and caution about the handling of sphygmomanometer body please follow the description of the instruction manual of sphygmomanometer This monitor has built in wireless equipment with construction desig...

Страница 46: ...uch as microwave ovens or in location with many obstructions or other environment subject to weaker radio wave signals There is a possibility that the wireless connection may break up frequently the c...

Страница 47: ...e manual of the receiver device to switch it to the state that a pairing is possible When pairing this monitor place it as close as possible to the receiver device to be paired with Hold down the SELE...

Страница 48: ...performed Ensure F20 is not set to OFF If the receiver device cannot receive measurement data try pairing again The communication distance between this monitor and the receiver device is dependent on...

Страница 49: ...Pairing described above Unpairing Devices can be unpaired Enter the Bluetooth utility mode Press the START STOP button with un in the systolic display section and PAr in the diastolic display section...

Страница 50: ...erations or at the start of first measurement When the setting function F20 is off the above clock synchronization is not performed 6 Transmission specifications Item Specifications Standard Bluetooth...

Страница 51: ...rcury sphygmomanometer or aneroid gauge Perform the inspection only as described below or the setting values and function settings may be changed After inspection check that the air connector plug is...

Страница 52: ...played 4 Using the rubber pump apply the pressures listed below Compare and check the pressures of the blood pressure monitor and the inspection equipment No Pressure setting Instrumental error A B st...

Страница 53: ...e main body and the arm cuff cover regularly When disinfecting them wipe them gently by using the gauze or dampened cloth with local antiseptic solution then wipe the moisture off the surface by using...

Страница 54: ...the power off and wait until the printer head has cooled completely The printer head gets very hot and may cause burns Some printer parts have sharp edges Take great care when handling them to avoid i...

Страница 55: ...ng out secure the printer paper by closing the printer cover until you hear a click If the cover is not completely closed a paper jam may occur Note When cleaning the printer head be careful of static...

Страница 56: ...d correctly into the connector Check the power cable for damage exposed core wires disconnection Check that the electrical outlet is properly grounded and supplies the specified voltage and frequency...

Страница 57: ...crew 2 Slide the front frame down and then pull forward 3 Loosen the screws armrest securing screws on the rear side and remove the screws 4 Lift the armrest and pull back 5 Slide the rear frame down...

Страница 58: ...teps used to remove reattach the rear and front frames return the armrest to its original position then replace the armrest securing screws 2 and front frame screw 1 Note The arm cuff cover is consuma...

Страница 59: ...tion display Press the START STOP button to reset the count 12 5 2 Printing the count graph To print the count graph Press the COUNT button While the number of measurements is displayed press the STAR...

Страница 60: ...up settings and other data Before disposing of the main unit remove the lithium battery and dispose of it according to local regulations Product name Model name Structure name Material Box Cardboard...

Страница 61: ...may not be possible with patients with arrhythmia or a weak pulse No printing The printer paper is not installed is displayed See 9 1 Installing the printer paper to install a new roll of printer pape...

Страница 62: ...patient was detected during measurement External vibrations may have been applied to the air system of the cuff or inside the monitor or an obstruction may have been mistakenly detected Check the pat...

Страница 63: ...ressure measurement Confirm the following conditions The cuff is not loose no thick clothing on the arm the patient remains still and no external vibrations on the cuff And try measurement again Pulse...

Страница 64: ...lem continues stop using the monitor immediately Restart the power A memory error was detected inside the monitor Restart the power If the problem continues stop using the monitor immediately There ma...

Страница 65: ...measure every day at the same time to learn your average blood pressure and blood pressure trends This blood pressure information will be important when visiting a doctor Consult with a doctor to dete...

Страница 66: ...ize of original bitmap patterns is as shown below Windows monochrome bitmap Create the bitmap data of the abovementioned size with a file name Logo bmp and save it in the root folder of the SD card No...

Страница 67: ...sfer mode 3 Insert the SD card containing the bitmap file Logo bmp saved in 15 1 Size of original bitmap patterns into the SD socket Press the START STOP button to start data transfer After transfer r...

Страница 68: ...tact 2 kV 4 kV 8 kV 15 kV air Radiated RF EM fields IEC 61000 4 3 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Proximity fields from RF wireless communications equipment IEC 61000 4 3 See table Test specifica...

Страница 69: ...nt Test frequency MHz Band MHz Service Modulation Maximum power W Distance m IMMUNITYTESTLEVEL V m 385 380 390 TETRA400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviatio...

Страница 70: ......

Страница 71: ......

Отзывы: