![A&D TM-2564G Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/aandd/tm-2564g/tm-2564g_instruction-manual_3099194040.webp)
29
5. Press the
or
key to change the value.
6. The changed value is confirmed when:
Moved to the next menu or sub menu.
The HOME key is pressed to return to the home
display.
Note
•
In the setting mode display, after about 5 minutes of inoperation, the home display will be indicated.
•
When setting the clock, note the following.
The time in the clock menu indicates that of when the clock menu is displayed, and it remains the same
as long as the clock menu is displayed. The actual clock of the instrument continues normal operation.
Changing any item of the clock menu is reflected and resets the clock.
•
When an alarm occurs, the display will return to the home display.
Setting items
Menu
Sub menu
Setting range
Unit
Description
Upper
OFF, 50-250 (by 10), 200
SYS
Lower
OFF, 30-200 (by 10), 70
Upper
OFF, 30-200 (by 10), 150
DIA
Lower
OFF, 10-200 (by 10), 40
mmHg
Upper
OFF, 50-220 (by 10), 170
HR/PR
Lower
OFF, 30-150 (by 10), 40
bpm
Upper OFF, 75-99 (by 1)
SpO
2
Lower
OFF, 50-99 (by 1), 85
%
Upper OFF, 30-44 (by 0.5)
Limits
Temp
Lower OFF, 26-40 (by 0.5)
°C
Sets the limit values for alarm.
Note
When the lower limit value for
SpO
2
measurement has been
set to be less than 85%, it will
be reset to the default setting
upon power-on.
Cycle
OFF, CON, 2.5, 3, 5, 10,
15, 20, 30, 60, 90, 120
min
Sets the cycles for interval
measurement.
Pressurization AUTO, 100-260 (by 20)
mmHg
Sets the pressurization value for
blood pressure measurement.
"AUTO": Detects the pulse at
the start of pressurization to
pressurize properly.
During interval measurements,
"AUTO" is used.
Patient Adult, Neonate
⎯
"Adult": Other than neonates
NIBP
Notice buzzer
OFF, ON
⎯
"ON": Sounds a buzzer 15
seconds before starting an
interval measurement.
Underlined items: Default settings
Содержание TM-2564G
Страница 1: ...TM 2560G TM 2564G TM 2560GP TM 2564GP VITAL SENSOR GRAPHIC MODEL INSTRUCTION MANUAL WM PD4000753 ...
Страница 11: ...ix ...
Страница 17: ...6 3 Part Names 3 1 TM 2560G TM 2564G ...
Страница 18: ...7 3 2 TM 2560GP TM 2564GP ...
Страница 62: ...51 Appendix B External Dimensions TM 2560G TM 2564G TM 2560GP TM 2564GP Unit mm ...