21
5-3 Trend Display
When the home display is being shown, press the TREND key (
key) to indicate the trend display of the latest 24
hours.
Note
•
When the trend display is being shown, press the HOME key to return to the home display.
•
In the trend display, after about 5 minutes of inoperation, the latest page will be indicated.
Name Displayed
item
Description
1
Number of data display
Number of data
•
Displays the number of data stored in memory.
2
Time scaling
Trend time width
•
Displays the time width of the trend currently displayed.
3
Page number
display
Page number
•
Displays the page number/total page to indicate where
the displayed data is among the whole data.
•
Displays blood pressure data in bar graph in the range
of 0 to 200 mmHg.
4 Scale
display
•
Displays SpO
2
data using in the range of 60 to 100%.
•
Displays heart rate/pulse rate data using
in the
range of 0 to 200 bpm.
5 Graph
display
Measured values
•
Displays temperature data using in the range of 25 to
45°C.
6
Trend end time
display
7
Trend start time
display
Time scaling of
the trend display
•
Displays the start and end time of the trend.
8
Scroll bar display
Scroll bar
•
Displays the bar graph to indicate where the displayed
trend is in the 24-hour scale.
Time scaling
•
Press the
/
keys to change the time scale.
(1, 2, 3, 6, 12, 24 hours).
9
Key operation
display
Scrolling
•
Press the
/
keys to scroll up or down the
pages.