34
6-7 Measuring the SpO
2
1. Turn on the POWER switch to start the SpO
2
measurement.
Note
•
In SpO
2
measurement, the function of automatically compensating the LED to obtain suitable emission
for the patient works. Thus, it may take approximately 30 seconds to display the measured value.
•
Do not apply the sensor on the same arm with the cuff attached. SpO
2
may not be measured correctly
during blood pressure measurement.
6-8 Measuring the Temperature (Optional)
1. Set the temp. probe to be used.
2. The instrument starts to receive the signal from the wireless temperature probe. Refer to the
instruction manuals of the temperature probe and the temperature receiver unit for details.
Note
•
The communication status is displayed by the antenna mark.
6-9 Measuring the Heart Rate/Pulse Rate
The heart rate and pulse rate are measured in three ways, using the instrument.
The values are displayed according to the order of precedence of ECG, SpO
2
and blood pressure
The heart mark starts to flash and the sound is generated to synchronize the pulse detection.
Note
•
The heart rate/pulse rate is displayed in different colors to indicate what measures the value.
Green = ECG , Blue = SpO
2
, Yellow = blood pressure
•
The synchronized sound can be turned off by the volume setting.
•
The heart rate/pulse rate measurement can be specified in the function setting. Refer to "6-14 Function
Setting".