43
8.4.
Deleting Measurement Data
Purpose of operation and explanation of function
Measurement data is deleted but settings are not deleted.
The initial settings can be configured using the following methods.
The method to use switches on the recorder.
The method to use
dedicated peripheral
that is connected to the
recorder using the USB cable.
If measurement data is deleted, it cannot be used again.
Backup data before deletion.
Delete measurement data of last patient before next patient
uses the recorder.
Several minutes may be needed to delete data.
Keep no-operation to delete data correctly.
Procedure of operation using switches
Step 1. If the indication of the display is hidden, press the START/STOP
or EVENT switch to return to the display of waiting mode.
Step 2. If the
mark is shown, press and hold the EVENT switch for
3 seconds or more to suspend A-BPM.
The
mark turns off.
Step 3. While holding the START/STOP switch, press and hold the
EVENT switch for
9
seconds or more.
DataClear will be displayed on the OLED display.
Step 4. Press and hold the START/STOP switch for
3
seconds or more.
Deletion of data will begin. Erasing will be displayed under
DataClear on the OLED display.
Several minutes may be needed to delete data.
Step 5. After the deletion, the recorder returns to waiting mode.