17
5.3.
Principal Switch Operations
5.3.1.
A-BPM Operations
To start or suspend A-BPM.
Step 1. Store the preset program (of start times and intervals) for A
-
BPM.
Step 2. Press and hold the EVENT switch for 3 seconds or more to
switch between the following states.
"
ON
"....... A
-
BPM is started and the
mark is shown.
Blood pressure measurements are performed in
accordance with preset A
-
BPM program.
"
OFF
"...... A
-
BPM is suspended and the
mark turns off.
Manual blood pressure measurement can be
performed by pressing the START switch, though.
To expand A-BPM interval time.
Step 1. Set the sleep mode to "
ON
" before the measurement.
Step 2. Start A-BPM by pressing and holding the EVENT switch for 3
seconds or more.
The
mark is shown.
Step 3. When the EVENT switch is pressed during A
-
BPM, the interval
time is doubled.
When the EVENT switch is pressed again, the interval time
returns to basic value.