35
The installation procedure is the same as for OP-03. Refer to “13.2.1. Installing OP-03”.
Set the function settings “
bp5
”, “
btpr
”, “
prt
”
,
“
5if
”
,
“
adr
” and “
aCk
” as
necessary.
Before using OP-04, the function setting “
5if
” must be set to specify whether RS-
422 or RS-485 is used.
To connect more than one scale to a computer, set a different address to each
scale using the function setting “
adr
”.
13.3.2. OP-04 Specifications
RS-422/485 Specifications
Transmission system EIA RS-422 / 485
Transmission form
Asynchronous, bi-directional, half-duplex
Data format
Baud rate: 2400, 4800, 9600 bps
Data bits:
7 bits + parity 1bit (EVEN / ODD)
or 8 bits (non parity)
Start bit:
1 bit
Stop bit:
1 bit
Code:
ASCII
Terminator:
C
R
L
F
(C
R
:0Dh, L
F
:0Ah)
LSB
0
1
2
3
4
5
MSB
6
Circuit diagram
Stop bit
A-B
1 (-2V~-5V)
0 (2V~15V)
Start bit
Data bits
Parity bit
1 SDA (RS-422 Out)
2 SDB (RS-422 Out)
3 RDA (RS-422 In / RS-485 In-Out)
4 RDB (RS-422 In / RS-485 In-Out)
5 TRM (Termination)
6 SG (Signal ground)
7 SHLD (Shield)
100
Ω
1 SDA (RS-422 Out)
2 SDB (RS-422 Out)
3 RDA (RS-422 In / RS-485 In-Out)
4 RDB (RS-422 In / RS-485 In-Out)
5 TRM (Termination)
6 SG (Signal ground)
7 SHLD (Shield)