60
Initialization (
Clr
)
Restores the following data to the default setting.
Function setting (Setting of “
5vtyp
” is not returned by this operation)
Calibration data
After initialization, check the viscosity value and perform calibration as necessary. (Refer to
8. VISCOSITY CALIBRATION").
1 In the standby mode, press and hold the
MODE key to enter the function setting
mode. "
5vtyp
" appears.
2 Press the MODE key to select "
Clr
".
3 Press the PRINT key to display "
Clr no
".
4 Press the START key to select "
Clr
go
".
5 Press the PRINT key to execute initialization.
After "
end
", the next item is displayed.
Initialization has completed.
6 Press the STOP key to return to the standby
mode.
Next item of
function setting
(Standby mode)
(Standby mode)
Press and hold