30
7 Turn the knob (Height direction) so as to
adjust the sample surface to the center of
the narrow part of the sensor plates.
Confirm that the sample surface is between
the upper and lower triangular marks.
Note
•
When using the 2 mL sample cup, raise
the protector.
•
Adjust the height of the sample surface
correctly, otherwise it may cause a
measurement error.
Model: SV-10A / SV-100A
1 Pour the sample into the cup until its surface is between the
level gauges.
- With the 45 mL sample cup, the lower level gauge indicates
35 mL, the upper level gauge indicates 45 mL.
- The 13 mL glass sample cup does not have level gauges.
The position approximately 25 mm above the cup bottom
indicates 13 mL.
2 Attach the cup on the table along the guides.
3 Confirm that the protector is in the position as shown in the
figure.
Raise the lever to release the sensor unit.
4 Pinch the grips, support the front side of the sensor unit and
gently lower the sensor plates above the sample surface.
5 Lower the lever to secure the sensor unit.
Sample cup (45 mL)
Guides
Protector
Sensor
plate
Knob
(Height direction)
Sensor protective
cover
Protector
Sample
surface
adjustment
position
Glass sample cup (13 mL)
Level gauges
Approx. 25cm