- 5 - 

CALIBRATION 

When to Calibrate 

 

  Calibration may be required when the scale is initially installed, if it is 

moved a substantial distance, or in accordance with local regulations.     
This is necessary because the weight of a mass in one location is 
not necessarily the same in another location. Also, with time and use, 
mechanical deviations can occur. 

The Position of the [UNITS] key for the Calibration 

 

  The (k)g model has no 

[UNITS] key, but the key 
for the calibration is 
located under the overlay. 

 
 
 

Calibration Using a Calibration Weight 

 

  Calibrate the scale with the following calibration weight. 

 

 

SK-1000WP: 1000g ± 0.1g  

SK-2000WP: 2000g ± 0.2g 

 

 

SK-5000WP: 5000g ± 0.5g 

SK-5001WP: 5000g ± 0.5g 

 

  SK-10KWP: 10kg 

± 

1g 

SK-20KWP: 20kg 

± 

2g 

  

SK-30KWP: 

30kg 

± 

2g 

 

 1.  Enter calibration mode. 

 

  Press the [ON/OFF] key to turn the power off.   

While pressing both the [RE-ZERO] and [UNITS] keys, press the 

[ON/OFF] key. Then, 

 Cal 

 will be displayed. 

 

 2. 

Zero 

calibration 

  

Press 

the 

[RE-ZERO] 

key to calibrate zero. 

 Cal0 

  will be displayed.   

 

  When you have proceeded to the display other than 

 Cal0 

, press 

the [ON/OFF] key to turn the power off and start over again. 

 

  Wait for the stable indicator

 

 to be displayed and press the 

[RE-ZERO] key. 

 Calf 

 will be displayed in a few seconds. If the 

zero calibration only is to be done, press the [UNITS] key. The scale 

will show 

 end 

  and automatically return to the weighing mode. 

 

 3. 

Span 

calibration 

 

  Place the weight at the center of the platform for accurate calibration.   

Wait for the stable indicator

 

 to be displayed and press the 

[RE-ZERO] key. The scale will show 

 end 

 and the scale will 

automatically return to the weighing mode.   

Содержание SK-1000WP

Страница 1: ... 0 ふ SK 1000WP SK 2000WP SK 5000WP SK 5001WP SK 10KWP SK 20KWP SK 30KWP 1WM PD4000461B ...

Страница 2: ...ter for the scale Other adapters may cause damage AC adapter is optional Internal service or adjustment to this product should be performed by a qualified person CAUTION Avoid installing the scale in direct sunlight which may cause discoloration or malfunctions Do not mix battery types or new and old batteries Replace with all new batteries at the same time If the scale is not to be used for a lon...

Страница 3: ...RE ZERO key Sets display to zero or subtracts weight of a container Four leveling feet Battery cover Batteries are not included AC adapter socket STABLE indicator Indicates when the reading is stable Polarity NET indicator Indicates when NET weight is displayed Tare function is used ZERO indicator Indicates when the scale zero is correct Weighing unit Battery compartment Battery cover lock Battery...

Страница 4: ...r Optional CAUTION When using an AC adapter the scale doesn t comply with IP65 The AC input requirement could be 100 120 220 230 or 240 volts 50 60Hz depending on the area where used so please verify that the adapter is the correct type Plug the AC adapter to the AC adapter socket inside the battery compartment Setting up the Scale CAUTION Avoid installing the scale in direct sunlight that may cau...

Страница 5: ...ale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Press the RE ZERO key to display 0 Verify the reading is 0 Place objects on the weighing platform to weigh When the reading becomes stable the stable indicator is displayed If you use a container for weighing Place an empty container on the platform Wa...

Страница 6: ...20KWP 20kg 2g SK 30KWP 30kg 2g 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO and UNITS keys press the ON OFF key Then Cal will be displayed 2 Zero calibration Press the RE ZERO key to calibrate zero Cal0 will be displayed When you have proceeded to the display other than Cal0 press the ON OFF key to turn the power off and start over again Wait ...

Страница 7: ...tion value 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO and UNITS keys press the ON OFF key Then Cal will be displayed 2 Select a digit to be changed Press the UNITS key The previously set value for example 979 8 will be displayed A digit after the comma will be changed by pressing the RE ZERO key Press the UNITS key to move the comma and sele...

Страница 8: ... kg 6 4 lb 3 2 kg 7 1 lb Calibration weight 1000g 0 1g 2000g 0 2g 5000g 0 5g 5000g 0 5g 10kg 1g 20kg 2g 30kg 2g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Manila 9 784 m s2 Athens 9 807 m s2 Melbourne 9 800 m s2 Auckland NZ 9 799 m s2 Mexico City 9 779 m s2 Bangkok 9 783 m s2 Milan 9 806 m s2 Birmingham 9 813 m s2 New York 9 802 m s2 B...

Страница 9: ... 8 World Map ...

Страница 10: ...tschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 대한민국 서울시 영등포구 여의도동 36 2 맨하탄 빌딩 8층 8th Floor Manhattan Bldg 36 2 Yoido dong Youngdeungpo ku Seoul KOR 전화 82 2 780 4101 팩스 82 2 782 4280 A D RUS CO LTD Компания ЭЙ энд ДИ РУС 121357 Российска...

Отзывы: