- 4 - 

 

WEIGHING 

Before Weighing 

Automatic Power-Off Function. 

 

  If the scale is left ON and the stable indicator is displayed, the automatic 

power-off function turns the power off after approximately five minutes 
(factory setting). To disable or enable the function, press the [ON/OFF] 
key while pressing the [RE-ZERO] key. The scale stores this setting. To 
verify this, press both the [RE-ZERO] and [UNITS] keys while in the 
weighing mode. (See "CALIBRATION" about the [UNITS] key.) 

 

 

poff0  

 Function disabled  

poff1  

 Function enabled 

Error Message 

 

 

    e    

 Overload 

 

 

 -----

 

  The scale zero is out of range 

 

 

   lb   

   Low battery 

Calibration May be Required Before Weighing 

 

  Read “CALIBRATION” first and if necessary, calibrate your scale for 

accurate weighing. 

 

Weighing Procedures 

 

1.  Turn on the scale. 

 

  Press the [ON/OFF] key to turn on the scale. Then the scale displays 

all segments for a few seconds and then 

  0 

  will be displayed. 

 

 2. 

Start 

weighing 

 

  If you do not use a container for weighing, 

 

  Press the [RE-ZERO] key to display 

  0 

. Verify the reading is 

  0 

 

  Place objects on the weighing platform to weigh. 

 

  When the reading becomes stable, the stable indicator

 

 is displayed. 

 

  If you use a container for weighing, 

 

  Place an empty container on the platform. 

 

  Wait for the stable indicator

 

 to be displayed and press the 

[RE-ZERO] key. Place the objects to be weighed in the container. When 

the reading becomes stable, the stable indicator

 

 is displayed. 

 

  Select the weighing unit, --- Only for the “(k)g-lb” model--- 

 

  If you use the (k)g-lb model, press the [UNITS] key to select 

weighing with either (k)g or lb as the unit. 

 

  Normally scales power up in lb when shipped from the factory. To 

power up in (k)g, turn the scale off. Press and hold the [UNITS] key, 

then press the [ON/OFF] key. The scale will start with (k)g. To return 

to the original setting, repeat this instruction. 

Содержание SK-1000WP

Страница 1: ... 0 ふ SK 1000WP SK 2000WP SK 5000WP SK 5001WP SK 10KWP SK 20KWP SK 30KWP 1WM PD4000461B ...

Страница 2: ...ter for the scale Other adapters may cause damage AC adapter is optional Internal service or adjustment to this product should be performed by a qualified person CAUTION Avoid installing the scale in direct sunlight which may cause discoloration or malfunctions Do not mix battery types or new and old batteries Replace with all new batteries at the same time If the scale is not to be used for a lon...

Страница 3: ...RE ZERO key Sets display to zero or subtracts weight of a container Four leveling feet Battery cover Batteries are not included AC adapter socket STABLE indicator Indicates when the reading is stable Polarity NET indicator Indicates when NET weight is displayed Tare function is used ZERO indicator Indicates when the scale zero is correct Weighing unit Battery compartment Battery cover lock Battery...

Страница 4: ...r Optional CAUTION When using an AC adapter the scale doesn t comply with IP65 The AC input requirement could be 100 120 220 230 or 240 volts 50 60Hz depending on the area where used so please verify that the adapter is the correct type Plug the AC adapter to the AC adapter socket inside the battery compartment Setting up the Scale CAUTION Avoid installing the scale in direct sunlight that may cau...

Страница 5: ...ale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Press the RE ZERO key to display 0 Verify the reading is 0 Place objects on the weighing platform to weigh When the reading becomes stable the stable indicator is displayed If you use a container for weighing Place an empty container on the platform Wa...

Страница 6: ...20KWP 20kg 2g SK 30KWP 30kg 2g 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO and UNITS keys press the ON OFF key Then Cal will be displayed 2 Zero calibration Press the RE ZERO key to calibrate zero Cal0 will be displayed When you have proceeded to the display other than Cal0 press the ON OFF key to turn the power off and start over again Wait ...

Страница 7: ...tion value 1 Enter calibration mode Press the ON OFF key to turn the power off While pressing both the RE ZERO and UNITS keys press the ON OFF key Then Cal will be displayed 2 Select a digit to be changed Press the UNITS key The previously set value for example 979 8 will be displayed A digit after the comma will be changed by pressing the RE ZERO key Press the UNITS key to move the comma and sele...

Страница 8: ... kg 6 4 lb 3 2 kg 7 1 lb Calibration weight 1000g 0 1g 2000g 0 2g 5000g 0 5g 5000g 0 5g 10kg 1g 20kg 2g 30kg 2g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Manila 9 784 m s2 Athens 9 807 m s2 Melbourne 9 800 m s2 Auckland NZ 9 799 m s2 Mexico City 9 779 m s2 Bangkok 9 783 m s2 Milan 9 806 m s2 Birmingham 9 813 m s2 New York 9 802 m s2 B...

Страница 9: ... 8 World Map ...

Страница 10: ...tschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 대한민국 서울시 영등포구 여의도동 36 2 맨하탄 빌딩 8층 8th Floor Manhattan Bldg 36 2 Yoido dong Youngdeungpo ku Seoul KOR 전화 82 2 780 4101 팩스 82 2 782 4280 A D RUS CO LTD Компания ЭЙ энд ДИ РУС 121357 Российска...

Отзывы: