A&D SB-30 Скачать руководство пользователя страница 9

SB-30/70/150 Operation Manual

9

Setup Menu Mode / 

Continued

SETUP MENU DESCRIPTIONS

NAME/CODE

DESCRIPTION

CODE/VALUE

F1

Display Mode

Selects display mode for the scale.

"0" = Non-NTEP mode

"1" = NTEP mode

0

1

√√√√

F6

Zero Range

Selects the range within which the scale may be zeroed. Note that
the scale must be in standstill to zero the scale.

100%

√√√√

1.9%

F9

Calib. Unit

Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"0" = primary unit is lb.                       "1" = primary unit is in kg.

0

√√√√

1

F11

Display Check

Actuates the function which illuminates all digit segments, decimal
points, and LCD annunciators in a test sequence. Pressing the

ZERO

 key to scroll down one level begins the test sequence.

Press 

ZERO

 key

to begin sequence

F12

Calibration

Places scale into the calibration routine. Scrolling down with the

ZERO

 key one level begins the procedure.

Press 

ZERO

 key

to begin sequence

F13

Factory Reset

This sub-menu will reset all parameters in the “F” and “A” menus to
the default settings. USE WITH CAUTION!!!!!

Press the 

ZERO

key twice to exe-
cute.

Calibration

ON/OFF

SWITCH

DC POWER

JACK

RS-232

PORT

REMOTE

DISPLAY

ID TAG

CALIBRATION

SWITCH COVER

CONNECTOR

(SB-30 & 70 ONLY)

CALIBRATION MODE KEY FUNCTIONS

ZERO

GROSS

NET

TARE

PRINT

SET

lb

kg

The scale can be calibrated by following the procedure shown below. The minimum test weight
and the recommended test weight depends on the scale’s capacity.

For the SB-30, the minimum test weight is 0.3 lb (0.15 kg) , and the recommended test weight is
20 lb (10 kg). For the SB-70, the minimum test weight is 0.7 lb (0.35 kg) , and the recommended
test weight is 47 lb (23 kg). For the SB-150, the minimum test weight is 1.5 lb (0.7 kg) , and the
recommended test weight is 100 lb (50 kg).

Содержание SB-30

Страница 1: ...S SB B S Se er ri ie es s D Di ig gi it ta al l B Be en nc ch h S Sc ca al le es s Instruction Manual Revision 1 3 April 5 2000 SB 30 SB 70 SB 150...

Страница 2: ...150 lb 70 kg capaci ties It comes standard with a bright LED screen for easy readout a removable stain less steel platform and an RS 232C serial communication port These scales can be configured to d...

Страница 3: ...e rocker switch to the ON 1 position The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT REMOTE DISPLAY ID TAG CALIBRATION SWITCH COVER SB 30 70 ONLY CONNECTOR SB 30 70 150 Back Pan...

Страница 4: ...never the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the...

Страница 5: ...n NET GROSS Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross...

Страница 6: ...wer off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc t...

Страница 7: ...rial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 A D NCI CAS Format 3 Transcell Technology...

Страница 8: ...tional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO key to begin 0 1 100 1 9 To pl...

Страница 9: ...ation Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F13 Factory Reset This sub menu will reset all paramete...

Страница 10: ...ged by pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then...

Страница 11: ...x 0 05 lb 0 70 x 0 02 kg CONSTRUCTION Base Steel and ABS Platform Stainless Steel Feet Non skid Hard Rubber DISPLAY 6 Digit 0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacit...

Страница 12: ...lect the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitting p...

Страница 13: ...ary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter installed Inst...

Страница 14: ...uch goods by Seller discloses that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing...

Отзывы: