HV/HW-CEP Series
Page 17
and lighting up
Blinking
M+
Blinking
M+
D
isplay and Symbols
Description
A
Calibration error
The calibration mass is too heavy.
Check the base unit and the weighing pan.
A
A
A
A
Does not display zero when the scale is turned on.
Remove anything that is on the weighing pan.
Perform zero point calibration.
Or
The weight value is unstable due to drift or vibration
when the scale is turned on.
A breeze or vibration may be affecting the measurement.
Check around the weighing pan.
A
A
A
A
Accumulated data count
A
A
Total mass value of the accumulated data
A
A
e.g. CAP. MAX. 3/6/15kg d=1/2/5g
The weighing range and measurable minimum mass.
Example: Displays the weight value by 5 g up to 15 kg.
Displays the weight value by 2 g up to 6 kg.
Displays the weight value by 1 g up to 3 kg.
A
A
7.2.
Keys
Display and Symbols
Description
A
ON/OFF key
The scale is in standby status when power is connected using
the AC adapter.
A
A
A
ZERO key
The
ZERO
key will zero the scale if the weight value is within
±2% of the weighing capacity (kg) around the power-on zero
point. The zero point indicator turns on (Zero operation).
If the scale is in tare in this time, the tare value is cleared.
When accumulation is displayed, the accumulation is cleared.
A
A
A
A
TARE key
The
TARE
key will tare the scale and subtract the weight to
zero as a tare (container) weight when the weight is a plus
stable value. In this case the zero point indicator and net
indicator turn on. (TARE operation)
When the tare is removed while the tare operation is in
progress and the scale returns to the zero point, the zero point
indicator and net indicator turn on.
In this case the displayed tare value will be negative.
Note The tare reduces the weighing range.
A
A
A
Adds to the accumulated data
.