22
7.3.
Voltage Output Selection
The range of output voltage can be changed using the slide switch on the panel.
Factory settings is "0V ".
"0V " (0 to 1 V)
: Zero 0.000 V, full scale 1.000 V
"0.2V " (0.2 to 1 V)
: Zero 0.200 V, full scale 1.000 V
Panel External View (Example: The GXM-06)
3 Analog GND
Analog output connector
4 Analog voltage output
Volume for fine adjustment
2
1
Output range
7.4.
Voltage Output Fine Adjustment
The voltage output has been adjusted at the
factory before shipment.
Using the ZERO volume and SPAN for fine
adjustment and a voltmeter, voltage output can
be finely adjusted.
Procedure
Step 1 Turn off the display using the ON:OFF key.
Voltage output becomes zero point status.
Adjust the ZERO volume so that the voltmeter indicates
0.000 V (when the slide switch is set to "0V ").
Adjust the ZERO volume so that the voltmeter indicates
0.200 V (when the slide switch is set to "0.2V ").
Step 2 While pressing and holding the PRINT and
SAMPLE keys, press the ON:OFF key to enter the
initializing mode. Then
1
V is output. Adjust the SPAN
volume so that the voltmeter indicates 1.000 V
Step 3 Repeat step 1 and step 2 until the correct voltage output is obtained.
GXM-06 RS-232C
GXM-06 RS-232C
The display to output 0 V (0.2 V).
The display to output 1 V.